72 Doc. 065289-03 10/12
4.5 Pump Does Not Start
No power (front Power LED indicator is not lit)
Check that the power cord is plugged in.
Check the main power fuses and replace, if needed (see Section 5.18
).
No communication between Dionex ICS-1100 and Chromeleon
Verify that the USB cable is connected correctly. For connection and setup
information, refer to the Dionex ICS-1100 installation instructions.
Pump is turned off
Turn on the pump from the Dionex ICS-1100 Control panel in Chromeleon.
Flow rate is set to 0
Set the flow rate from the Dionex ICS-1100 Control panel in Chromeleon.
4.6 No Flow
Pump waste or priming valve open (see Figure 4-2)
Close the valves by turning the knobs clockwise until fingertight. DO NOT
OVERTIGHTEN! Overtightening may damage the valve and the pump
head.
Figure 4-2. Waste and Priming Valves
Waste Valve
Priming Valve