User guide

34 Operation
3.11 Setting the Heaters
The only available Heater in the dispensing system is the needle heater. Heater temperature is set in the
FmXP software. The heater controller monitors the Resistance Temperature Detector (RTD), compares it
to the heat parameters set in Fluidmove, and regulates the voltage going to the heating element to
maintain the set temperature.
3.11.1 Needle Heater
CAUTION! The operator can turn the heaters ON and OFF and monitor their function.
However, only authorized personnel should set heater parameters.
NOTE Contact your applications engineer or Nordson ASYMTEK Technical Support for
information about setting heater parameters for your particular application.
To setup the needle heater in Fluidmove:
1. In the Fluidmove Main Window, click on Run a Program .
2. In the Production Window (Figure 3-14), click on Setup and in the list box select
Setup Heater 1.
Figure 3-14 Fluidmove Production Window
The Heater Control Window (Figure 3-15) opens.
3. In the Heater Control Window, locate the channel (loop) for the desired heater.