Technical data

4. Check the closing caps in the backup boxes
In order to carry out this part of the testing
successfully, you need to make sure that all the 90°
bends of the backup boxes are provided with the
taps for the airtight closure of the system.
26
Location of
losses
Central vacuum systems
6. Switch on the testing vacuum cleaner
Wait a few seconds so that the system creates the
depression needed for the measurement. The
vacuum gauge will indicate the exact depression
value once the pointer has come to a standstill.
7. Measure the depression at the inlet valve
Check that the depression value measured at the
inlet valve corresponds with that at the inlet of the
testing vacuum cleaner. If the tubes and the fittings
have been glued in a perfect way, the values will
coincide.
Please notice: if the measured values do not
correspond, there is an air infiltration in the tubing.
1. Locating the origin of the whistles
Follow systematically the pipe line and check if it is
damaged. At the same time, listen if there are any
whistles. If this check does not provide any
successful result, use the stethoscope (part of
Disan’s testing kit) to listen systematically along the
tubing for whistles; in particular where fittings and
tubes have been glued. Even the minimum loss
produces irritating noises!
2. Repair the tubing
If necessary, replace the tube or the fittings so that
the tubing is perfectly tight.
Repeat the testing of the system.
After carrying out the test successfully, Disan
suggests giving to the purchaser or to the chief of
works a copy of the certificate (see second-last
page of this handbook) and of the tubing distribution
in order to avoid disputes or future liabilities.
5. Insertion of the vacuum gauge
Insert the vacuum gauge directly into the inlet
socket mounted on the vacuum cleaner.
Z 203
Z 204