FOREHEAD- AND-EAR THERMOMETER DX 6635 GB USER MANUAL • Precise measurement of body temperature in the ear and on forehead • Audio output of readings in German and English • Integrated light for nighttime usage • Digital display with blue back lighting • Additional function: Display shows time and room temperature (in standby mode) • Date, month and year display • Previous body temperature measurements saved • Practical stand for storage and safekeeping • Includes 2 x 1.
GB 1.0 2.0 3.0 3.1 3.2 3.3 3.4 3.5 3.6 3.7 3.8 3.9 4.0 4.1 4.2 4.3 4.4 4.5 5.0 6.0 7.0 8.0 9.0 10.0 11.0 12.0 13.0 14.0 15.0 16.0 17.0 18.0 TABLE OF CONTENTS What the symbols mean..............................................................3 Usage and function of the forehead-and-ear thermometer...3 Safety Information........................................................................4 General Safety Information........................................................
SYMBOLS AND FUNCTIONS 1.0 GB What the symbols mean The safety symbols shown in this manual are indications for the proper usage of the instrument and for the protection of your safety. The symbols stand for the following contents: Always read and respect the user manual. Caution/Risk: If used improperly, there is a risk of serious injury/damage. These directions must absolutely be adhered to.
SAFETY INFORMATION GB 3.0 Safety Information 3.1 General Safety Information 3.1.1 The forehead and ear thermometer DX 6635 is a sensitive medical products. It may not be subjected to shocks, blows or falls as these may cause a defect or destroy the product altogether. If such interference occurs, the measuring results may be faulty. If this is the case, contact our service center. (See page 27) 3.1.
SAFETY INFORMATION 3.2 GB The uses and environs for which the foreheadand-ear thermometer is suitable 3.2.1 Only use the forehead-and-ear thermometer for the designated intended purpose - the measuring of human body temperature. 3.3 The uses and environs for which the forehead-andear thermometer is not suitable 3.3.1 This instrument is not suitable or appropriate for medical or commercial use. It is intended for household/domestic use. 3.
GB 3.5 SAFETY INFORMATION The types of measurement for which the forehead-and-ear thermometer is not suitable 3.5.1 The forehead-and-ear thermometer DX 6635 is not suitable for the following types of measurement: - Axillary measurements (under the arm) - Oral measurements (in the mouth) - Rectal measurement (in the anus) 3.6 Usage by children and young persons 3.7 Storage/maintenance of the instrument 3.6.1 Keep the forehead-and-ear thermometer out of reach of children.
SAFETY INFORMATION 3.8 Cleaning and care of the instrument 3.9 Disposal of the instrument GB 3.8.1 The forehead-and-ear thermometer will not tolerate humidity/damp, vibration or being struck. Do not hold the forehead-and-ear thermometer under running water; do not immerse it in water or other fluids. 3.8.2 The forehead-and-ear thermometer is sensitive to heat, and must not be exposed to direct sunlight. Do not place the forehead-and-ear thermometer on any hot surfaces. 3.8.
GB QUESTIONS REGARDING Body TEMPERATURE 4.0 4.1 Questions regarding aural (ear) temperature What is aural temperature? 4.2 How is aural temperature measured correctly? 4.3 Why is the ear the best place to measure body temperature? The body temperature relates to the internal temperature of the body (core temperature). Because of this, the inner ear area can accurately reflect fluctuations in body temperature.
QUESTIONS REGARDING Body TEMPERATURE GB body temperature measurements. Normally, rectally measured body temperatures will be 0.3°C higher than body temperatures taken orally, while this area in turn gives values 0.3°C higher than those taken from readings in the armpit. In order to establish your personal normal body temperature, and obtain readings for family members, too, you can take body temperature readings at differing times of the day, over the course of 3 or 4 days, when you are in good health.
GB 5.0 PRODUCT CONTENTS/DESCRIPTION OF PARTS Product contents and description of instrument parts 1 2 3 9 11 4 1. 2. 3. 4. 5. 6. 7. 8. 9. 10. 11.
LCD DISPLAY INDICATION 6.0 GB LCD Display Indication 18 19 17 16 15 14 13 10 12 2 1 9 11 3 4 5 6 7 1. Display showing day of the week 2. AM/PM when in 12 h display mode 3. Battery is empty 4. Memory display when in memory call-up 5. AM/PM when in memory display 6. Y stands for Year 7. Year in clock mode, and hours in memory mode 8. Year in clock mode, and minutes in memory mode 9. Month and day symbol in memory mode 10. Month and day symbol in clock mode 11.
BASIC FUNCTIONS GB 7.0 Basic Functions 7.1.1 In standby mode, the current date, time and ambient temperature will be displayed. The date display range lies between 2001-2099 (Gregorian calendar), the time display can be set to be displayed in either 12 or 24-hour format. The starting time when instrument is reset is set to be 01.01.2008. An announcement of the time and the ambient temperature can be made by pressing the [TALK/OK] button.
TIME MODE 8.0 GB Time mode Step 1 - standard display The following data will be displayed when in time mode: Year, month, day, hour, minute and ambient temperature reading. The two flashing dots between the hour and minute displays indicate the passing of second-long intervals. The display of hours and minutes (see Fig. 1) alternates roughly every 5 seconds with the ambient temperature display (see Fig. 2).
Memory mode GB 9.0 Memory mode - calling up the measuring results D Press the [MEM/SET] button when in time display mode in order to be able to read off the saved temperature measurement readings. If no measurement reading has taken place, the display will appear as shown in illustration fig. D. If temperature measurements have taken place, then these will be displayed in turn by each successive pressing of the [MEM/SET] button (see fig. E).
SET-UP MODE GB 10.0 Set-up mode Step 1: In order to get into the set-up mode, you should hold down the button [MEM/SET] for around 2 seconds. Now the 12 or 24 hour setting will flash. If the [MEM/SET] button is now pressed down once again, you can set the display to be on either 12 or 24 hours.
GB Scope of supply/part data 11.0 Measuring temperature in the ear Fig. 1 Make sure, before taking each ear temperature reading, that the measuring sensor and the ear canal are clean, because otherwise the reading will be inaccurate. For ear temperature to be taken, the plug-in attachment (see page 10, Nr. 5, Fig. 1) must be taken off. Pull the ear gently back and insert the measuring probe (measuring sensor) along the ear canal, pointing towards the eardrum. Set the sliding switch (page 10, Nr. 5, Fig.
CHANGING THE BATTERY GB 13.0 Changing the battery A C B Step 1: Press down lightly on the cover of the battery compartment and slide it out to the right (see Fig. A). Now you can take out both of the AAA batteries and insert new batteries in their place. Pay attention to the correct polarity (+ = positive pole, - = negative pole). The correct position of the batteries is indicated on the bottom of the battery compartment.
Technische Störungen GB 14.0 Technical malfunctions Error No response Checklist Batteries flat? Countermeasure Insert new batteries. Batteries badly inserted? Insert batteries correctly. Battery symbol appears on the LED-Display “Lo“ shown on display (measurement result under 32.0°C) “Hi“ shown on display (measurement result over 42.9°C) “ErrE“ shown on display ErrU“ shown on display “ErrH“ shown on display Low Battery Insert new batteries.
Geräte- und Batterieentsorgung GB 15.0 Disposal of the batteries: Used batteries should not be placed in household waste. Dispose of them via your specialized electronics dealership or your public materials collection point. As a consumer, you a legally obliged to give back used batteries. These signs indicate a battery containing harmful substances: Pb = contains lead Hg = contains mercury Cd = contains cadmium.
GB TECHNICHAL DATA AND SPECIFICATIONS 17.
TECHNICHAL DATA AND SPECIFICATIONS Working air humidity: Fluctuations/tolerances: Storage temperature: Storage air humidity: CE0123: Lot number: Series number: 2009-05 GB 20 - 85% air humidity +/- 1 °C when measuring the ambient temperature -10 - 55° C 20 - 85% air humidity The device was certified in acc. with EU Guideline 93/42/EWG „Medical products“. Tested in acc. with EN ISO 60601-1-2:2001 acc. to EU Guideline on electromagnetic tolerance (EMV) 89/336/EWG.
GB TECHNICHAL DATA AND SPECIFICATIONS Guidelines and Manufacturer‘s Declaration - Electromagnetic Emissions The model DX6635 is designed for operation in an environment as stated below. The customer or user of the model DX6635 must ensure that it is operated in such an environment. Interference emissions measurements Conformity Electromagnetic environment - guideline HF emissions in acc. with CISPR 11 Group 1 The model DX6635 only uses HF energy for its internal function.
TECHNICHAL DATA AND SPECIFICATIONS GB Guidelines and Manufacturer‘s Declaration - Electromagnetic Jamming Resistance The model DX6635 is designed for operation in the electromagnetic environment stated below. The customer or user of the model DX6635 must ensure that it is operated in such an environment.
GB TECHNICHAL DATA AND SPECIFICATIONS Guidelines and Manufacturer‘s Declaration - Electromagnetic Jamming Resistance The model DX6635 is designed for operation in the electromagnetic environment stated below. The customer or user of the model must ensure that it is operated in such an environment.
TECHNICHAL DATA AND SPECIFICATIONS REMARK 1 GB At 80 MHz and 800 MHz, the higher value applies. REMARK 2 These guidelines may not be applicable in all situations. The spread of electromagnetic waves is influenced by the absorptions and reflections of buildings, objects and people. a It is theoretically not possible to exactly determine the field strength of stationary transmitters, e.g.
GB TECHNICHAL DATA AND SPECIFICATIONS REMARK 1 At 80 MHz and 800 MHz, the higher frequency range applies. REMARK 2 The ISM frequency bands (for industrial, scientific and medical applications) between 150 kHz and 80 MHz are 6.765 MHz to 6.795 MHz; 13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz and 40.66 MHz to 40.70 MHz. REMARK 3 The coincidence levels in the ISM frequency bands between 150 kHz and 80 MHz and in the frequency range from 80 MHz and 2.
WARRANTY GB 18.0 Warranty The forehead-and-ear thermometer you have purchased has been developed and manufactured with utmost care. The warranty period totals 36 months after the purchase date and covers both, material and manufacturing deficiencies of the product. Please retain the receipt as proof of purchase, which will be required for the filing of warranty claims.