Manual

! !
! !
Fiber-Lite
®
Operation
All the MI-150 series illuminators are equipped with a front panel manual solid state intensity control.
Press the ON(1)/OFF(0) rocker switch on the front panel to ON(1). The illuminator will light and is ready for
service. The illuminator intensity is controlled by the rotary control located on the front panel of the illuminator.
The 0 position (when control is turned fully counter-clockwise) corresponds to the lowest illuminator intensity.
The 100 position (when control is turned fully clockwise) corresponds to the highest illuminator intensity.
NOTE: Continuous operation of the illuminator at the highest intensity level will yield rated lamp life.
Operating the illuminator at reduced intensity can result in significantly extended lamp life.
WARNING
Risk of electrical shock. Remove power plug before lamp
replacement and wait for hot lamp to cool.
Lamp Replacement
Turn the illuminator intensity control fully counterclockwise (the 0 position) and run the illuminator with the fan for
several minutes.
Wait until the nosepiece is cool to the touch.
Press the ON(1)/OFF(0) rocker switch to the OFF(0) position.
Remove the AC power cord from the AC power receptacle.
Open the lamp access door by depressing the small clip toward the rear of
the illuminator with your thumb or finger. Once the clip disengages from the
illuminator housing the lamp access will spring open several inches. The
o
lamp access door will open slightly more than 90 to allow easy access to
the lamp and lamp socket.
Do not pry open the door with a tool. This may result in damage to the
illuminator housing or the lamp access door.
Check the lamp assembly to verify that the lamp and socket are cool before
proceeding. Caution: The lamp runs at very high temperatures and
contact with a hot lamp may result in severe injury.
Pull the lamp eject lever to remove the lamp from the socket.
Lift and remove the lamp from the lamp socket by grasping the rear of the
lamp adjacent to the lamp
power pins.
Remove the lamp from the socket.
Discard the old lamp.
Insert the replacement lamp into the lamp socket by aligning the lamp power pins with the slots in the lamp socket.
CAUTION: Do not touch the interior of the lamp reflector, the lamp envelope or the lamp pins with your
fingers. Touching the interior of the lamp reflector, the lamp envelope or the lamp pins will result in
significant shortening of the lamp life. Handle the lamp only by the exterior of the reflector or the area
adjacent to the pins.
Push the lamp gently into the socket to seat the lamp in the socket and lampholder.
Rotate the lamp access door toward the illuminator housing and gently press the lamp access door to cause the door
clip to seat into the housing and secure the door.
Reattach AC line cord and the illuminator is ready for service.
WARNING
Replace the fuse with the correctly rated fuse as listed on the label
on the back of the illuminator. Use of an improper fuse can create a
FAX: 508-682-2500
13.
1.
2.
3.
4.
6.
5.
7.
9.
10.
12.
11.
8.
Filter Option:
MI-150 illuminators ordered with the filter option are shipped with the IR filter and the filter holder installed at the factory.
The filter holder s located behind the illuminator nosepiece, in front of the lamp. (see diagram on page 2)
To use color filters (not included), place color filter into the filter holder behind the IR filter such that the IR filter is located
between the lamp and the color filter.
WARNING: Do not use color filters without an IR filter in place. Filters become HOT with use! Allow sufficient time (10-15
minutes) for the filters to cool before handling.
2
Fuse Replacement
Press the ON(1)/OFF(0) switch to the OFF(0) position.
Remove the AC line cord from the AC power receptacle.
Remove the AC line cord from the power entry module at the rear of the illuminator. The fuse drawer is part
of the power entry module. The drawer is located directly beneath where the AC line cord plugs in.
For 115 VAC illuminators: Pull out the fuse drawer. The AC power fuse is located in the receptacle with
metal clips at both ends of the fuse.
Remove the blown fuse and discard. The second fuse is the spare.
Place the replacement fuse into the metal clips in the fuse drawer. The fuse will work in either orientation.
Push the fuse drawer into the illuminator until it "clicks" into position.
For 230 VAC illuminators: Push down the center latch until it locks in the down position. Remove the fuse
drawer. Two fuses are required.
Remove the blown fuse(s) from the drawer and discard.
Place the replacement fuse(s) into the fuse holder by inserting either end of the fuse into the fuse holder.
Push the fuse drawer until it "clicks" into position.
Attach the AC line cord to power entry module at the rear of the illuminator. The illuminator is now ready for
service.
Replacement Parts
Part No. Description
02-015850-0000 Lamp, EKE, 21V, 150 Watt
02-015852-0000 Lamp, EJV, 21V, 150 Watt
02-015855-0000 Lamp, EJA, 21 V, 150 Watt
02-015320-1300 Fuse, 3A, 5 x 20, 250 V, slow blow (115 VAC)
02-015320-3200 Fuse, 2A, 5 x 20, 250 V, slow blow(230 VAC)
Only above parts are replaceable. Return illuminator to factory for warranty service.
Attempts to replace other parts voids warranty.
Performance Statement
Dolan-Jenner Industries, Inc. (DJI) recognizes that its illuminator products may be used under an almost unlimited
variety of conditions. As such, we are prepared to assist the customer in the selection and application of any of these
products. This includes application engineering, sample testing and other means as determined by DJI.
Where DJI has made specific recommendations for its products, systems, or detection techniques (based on complete
and detailed information furnished by the customer) we will extend every effort to assure that the customer is satisfied with
the performance of our products. Continual development and improvement of DJI products may require changes in
details that do not coincide with descriptions or illustrations shown. All fiber optic cable diameters are nominal.
Three Year Warranty on Light Sources
Dolan-Jenner Industries, Inc. (DJI) warrants its products to be free from defective material and workmanship. Any light
source or parts thereof which are determined by DJI to be defective within 3 years (average product life cycle) from
shipment date will be replaced or repaired at our option. This policy is effective Nov. 1, 1993 and is not retroactive. All
fiber optics are warranted for one year. This warranty specifically excludes both incandescent and quartz-halogen lamps,
and optical filters.
Any products which, in our opinion, have been subjected to misuse, incorrect wiring, or where installation procedures
are not in accordance with the instruction manual, are excluded from this warranty. Nor does this warranty extend to
products on which repairs or alterations have been made outside the factory, or on which the identification or serial
number has been altered or to accessories not of our manufacture.
Our obligation with respect to products or parts covered by this warranty shall be limited to repair or replacement,
F.O.B., Boxborough, Massachusetts. In no event shall DJI be held liable for consequential or special damages, or for
transportation, installation, adjustment, or other expenses which may arise in connection with such products or parts. This
warranty is in lieu of all other statements or warranties or guarantees, written or implied, by DJI or its authorized
representatives.
Important: Please contact factory for a return authorization number prior to shipping merchandise to factory.
Fiber-Lite
®
POWER
ENTRY
MODULE
FUSE
DRAWER
1.
2.
3.
4.
6.
5.
7.
9.
10.
12.
11.
8.
LAMP
FILTER
HOLDER
LEVER
Pull back to eject lamp
SIDE DOOR
3