User manual

16
12.
When fresh vaccines are stored in the ILR, the cabinet
temperature may go up initially but should fall and
come to the required range after few hours.
If it is seen that the temperature does not come down
to the required level and remains steady at a higher
degree, the thermostat may have to be adjusted to a
higher setting.
After adjustment always observe till the cabinet
temperature comes down to the required level and
remain steady there.
13.
The cabinet temperature may vary depending upon
various factors, such as,
the ambient temperature
amount of vaccine preserved in the ILR ยท
frequency of lid opening
circulation of air around the ILR, etc.etc.
If such variations are observed to be continued for
longer time, the ILR temperature may be brought to
the required range, by adjusting the thermostat as
described earlier.
TO USE ILR PRESENTLY RUNNING AS
`REFRIGERATOR' TO WORK AS `FREEZER'
(Change-over from REFRIGERATOR to FREEZER
function)
1) Take out the stored vaccines, which are not to be
preserved at freezer temperature, from the ILR
and preserve them in alternative storage.
2) Follow actions detailed in Chapter "To use as
FREEZER".
4. TO USE A REFRIGERATOR
1.
Install and prepare the ILR as detailed under
`Installation' and `Preparation' in the previous
chapter.
2.
The Refrigerator/Freezer function selector switch is
located at left upper portion in the compressor
compartment. This switch is accessible from outside
(fig. 27)
Normally the ILR is delivered with this switch set for
Refrigerator function.
Examine this switch and shift it to the RIGHT for
FREEZER function.
5. TO USE A FREEZER
Fig. 27.