DEFROSTING & CLEANING
PROCEDURE
1.
Switch OFF the power supply to the ILR and remove
the plug of its power supply cord from the wall socket.
2.
Prepare the temporary storage and transfer the
vaccines and preserve them properly with care.
3.
Take the frozen ice lining tubes out. Keep them close
together in a upright position, wrapping them together
with a thick blanket or similar insulating material.
This will keep the coldness of the tubes during
defrosting of the ILR.
4.
Take out the plastic floors from the bottom (see fig. 15
on page 1 1 ). Then remove the frozen ice packs from
the bottom of the ILR and keep them wrapped, too, to
preserve their coldness, similar to the lining tubes.
6. DEFROSTING & CLEANING