User manual

23
7.
Clean the ILR carefully as below,
a) Wash all parts inside the cabinet with warm water
and mild detergent.
b) Clean the lid and the lid seal (gasket) similarly After
the rubber seal (gasket) is wiped and dried, it should
be rubbed with unscented talcum powder (or French
chalk) specially on the hinge side.
c) Clean the plastic inner liner, floors and the outside
of the ILR, also with warm water and mild detergent.
8.
Allow the cleaned parts to dry completely.
9.
Reset the drain plug at its position at the bottom
10.
Replace the layer of the frozen ice packs on the
bottom, as before, wiping each ice pack before
placing.
Replace the plastic floors.
11.
Take the plastic ice lining tubes one by one, wipe
their outside with a clean dry cloth and put them
back carefully into their positions.
Examine the tubes for any leakage of water from
any portion, before putting them back to the ILR. A
defective tube will have to be either repaired or
replaced.
12.
Close the lid. Connect the power supply plug to the
wall socket and switch ON the ILR.
13.
Allow the ILR to run and observe the cabinet
temperature on the thermometer.
14.
Put back the vaccines into the ILR from the
temporary storage, only when the ILR has attained
safe recommended temperature range for storage
of the vaccines.
6. DEFROSTING & CLEANING