Product Manual

19
Error Messages:
Error Message: What to do or check:
“Er EP” The setup parameters loaded in
nonvolatile memory have become
corrupted.
Verify scale parameters and
calibrate.
“rg Er” The calibration zero is out of range.
Error is displayed after a ZERO
calibration attempt. Press zero to
clear this error.
Refer to the analog setup section for
additional information.
Pcb or load cell may need to be
replaced.
“Ldg 0” The scale is attempting to zero on
power-up.
This message will remain until the
scale is stable. Air currents or
vibration may be the cause. If
problem persists, the pcb or load
cell may be damaged.
NOTE: This message will not
appear if parameter Suo = no.
“ov-Ld” The scale is in overload. The load
on the scale platform exceeds the
scale capacity by more than 105%.
Remove excess weight from scale
platform.
If problem persists, recalibrate.
If problem still persists, the pcb or
load cell may need to be replaced.
“gS-oL” The scale is in gross overload. The
load exceeds the scale rating and
might result in damage to the scale.
Remove excess weight immediately.
If problem persists, recalibrate.
If problem still persists, the pcb or
load cell will need to be replaced.