4) Place full capacity (2lbs for the 2.2lb scale and 12lbs for the 13.2lb scale) on
the platter and press ZERO. Wait for the scale to count down from “7” to “0. If
the scale detects motion, the count will restart. When finished, the scale will
return to the weigh mode.
NOTE: If “SPnL” or “SPnH” appears on the display, the calibration span is out of
range. Verify that the calibration weight is correct and repeat the
calibration. Refer to the analog setup section for additional information.
5) If additional parameter changes are needed, refer to Chapter 4.
Analog Setup:
The following table shows the acceptable “Raw Counts” for no load and full load.
“Raw Counts” can be viewed through the last setup parameter. Refer to Section
4, Parameter Setup, for more information regarding parameter viewing.
If the scale’s raw counts are not within the ranges specified in the following table,
load cell may need to be replaced.
Scale Capacity Acceptable “Raw Counts”
for no load
Acceptable CHANGE in
“Raw Counts” for full load
2 lb 18,500 – 41,000 64,500 – 92,500
5 lb 14,500 – 35,250 110,000 – 143,000
10 lb 6000 – 24,000 120,500 – 153,000