Chapter 5 Calibration
1000-DCV DC Volts
Equipment Required:
1. Precision DC voltage source with a resolution to 0.1mV and
an accuracy of ±0.1%
2. Interconnecting copper wire from the DC source to the
indicator.
3. Trimmer adjusting tool.
STEP 1. Disconnect power from the unit and use copper
connecting leads, connect the output of the DC voltage source to
terminals 5 (+) and 6 (-) of the indicator according to the chart on
page 5.
STEP 2. Remove the front panel lens, reconnect power and allow
for a 5-minute warm-up.
STEP 3. Input 0.0 mV from the DC voltage source and adjust the
ZERO pot for a display of 0.0±1 count. (Refer to the figure on page
15 for the location of the zero and span pots).
STEP 4. Input 190.0 mV from the DC voltage source and adjust
the SPAN pot for a reading of 190.0±1 count.
STEP 5. Repeat steps 3 and 4 until no further adjustment is
needed.