User Manual

21
1000-FREQ Frequency
Equipment Required:
1. Precision AC voltage source with a resolution to 0.1mV and
an accuracy of ±0.1%
2. Interconnecting copper wire from the AC source to the
indicator.
3. Trimmer adjusting tool.
STEP 1. Disconnect power from the unit and using copper
connecting leads, connect the output of the AC Voltage source to
terminals 5 and 6 of the indicator according to the chart on page 5.
STEP 2. Remove the front panel lens and reconnect power and
allow for a 5-minute warm-up.
STEP 3. Input 100.0 mV at 60 Hz from the AC voltage source
and adjust the ZERO pot for a display reading of 60±1count.
STEP 4. Input 100.0 mV @400 Hz from the AC voltage source
and adjust the SPAN pot for a display reading of 400±1count.
STEP 5. Repeat steps 2 and 3 until no further adjustment is
needed.