User Manual

24
0 to 20mA
Equipment Required:
1. Precision DC current source with a resolution of 0.1ma DC
and an accuracy of 0.1%.
2. Interconnecting copper wire from the DC current source to the
indicator.
3. Trimmer adjusting tool.
STEP 1. Disconnect power from the unit and remove the front
panel lens.
STEP 2. Connect the DC current source to terminal 1 (+) and
terminal 2 (-) located on the rear of the unit. Reconnect power and
turn the unit on. Allow for a 5-minute warm-up.
STEP 3. Input 0.0mA from the DC current source and adjust the
ZERO pot for a display reading of 0.0 ± 1 count.
STEP 4. Input 20.0 mA from the DC current source and adjust
the SPAN pot for a reading of 100.0±1 count.
STEP 5. Repeat steps 3 and 4 until no further adjustment is
needed.