User Manual

25
1 to 5VDC
Equipment Required:
1. Precision DC voltage source with a resolution of 0.1mV DC
and an accuracy of 0.1%.
2. Interconnecting copper wire from the DC voltage source to the
indicator.
3. Trimmer adjusting tool.
STEP 1. Disconnect power from the unit and remove the front
panel lens.
STEP 2. Connect the DC voltage source to terminal 3 (+) and
terminal 2 (-) located on the rear of the unit. Reconnect power and
turn the unit on. Allow for a 5-minute warm-up.
STEP 3. Input 1.0VDC from the DC voltage source and adjust
the ZERO pot for a display reading of 0.0 ± 1 count.
STEP 4. Input 5.0VDC from the DC voltage source and adjust
the SPAN pot for a reading of 100.0±1 count.
STEP 5. Repeat steps 3 and 4 until no further adjustment is
needed.