5.2 Mounting the DORO Swivel Adaptor
Mount the Swivel Adaptor to the Transitional Member of the Base Unit:
1. Position the teeth of the starburst connection of the
Transitional Member and the teeth of the lower starburst
connection of the Swivel Adaptor in a manner that they
engage.
2. Fully tighten the lower locking screw of the Swivel Adaptor
by turning it clockwise. Note: the teeth of the starburst con-
nections must be fully engaged and seated properly.
5.3 Mounting of the DORO Skull Clamp
Mount the Skull Clamp (Item-No. 3003-00 or 1001.001) to the Swivel Adaptor
1. Position the teeth of the upper starburst connection of the
Swivel Adaptor and the teeth of the starburst connection of
the Skull Clamp in a manner that they engage.
2. Position the Skull Clamp as desired.
3. Fully tighten the upper locking screw of the Swivel Adaptor
by turning it clockwise. Note: the teeth of the starburst
connections must be fully engaged and seated properly.
Danger:
Check all fastening screws and all starburst locking mechanisms on the de-
vice and check the stability of the complete DORO system before and after
each clinical use. Make sure that the device is exactly mounted as described
in this manual.
5.4 Mounting of the DORO Skull Pins
This system requires three DORO Skull Pins of the same versions to be used. Please refer
to the instruction manual of the appropriate DORO Skull Pins.
Important:
QR3 only : The main starburst of the QR3 Skull Clamp (base) is designed for con-
necting the QR3 Skull Clamp to DORO Swivel Adaptors and other accessories.
When connecting non-DORO accessories with similar starburst connections, test
that the connection is tight and secure and the teeth of both starbursts are fully en-
gaged with no gap. A poor connection could be caused by a screw on the accessory
which is too long. These accessories must not be used with the QR3 Skull Clamp.