Instruction manual
The following 3 steps - - are required:
Mount the Base Unit and the Swivel Adaptor to the OR
Table
1. Mount the Base Unit to the OR Table (see Chapter Moun-
ting).
2. Mount the Swivel Adaptor to the Transitional Member of
the Base Unit (see Chapter Mounting).
3. For ease in mounting, we recommend that the Transitio-
nal Member of the Base Unit and the Swivel Adaptor are
loosely connected (not fully tightened) and the Swivel
Adaptor is hanging loose and pointing to the floor.
6. Use and Handling
6.1 Detailed description of application
NOTE:
The user decides the best approach to secure the patient to the Skull
Clamp. We recommend to secure the patient‘s head rst to the Skull
Clamp with Skull Pins and then to connect the Skull Clamp carefully by
means of the Swivel Adaptor to the Base Unit and the OR Table.
A B C
A
Secure the patient's head to the Skull Clamp
(Item-No. 3003-00 or 1001.001)
Insert the Skull Pins into the receptacles of the Skull Clamp
(refer to instruction manual Skull Pins).
Adjusting the required Skull Clamp width:
1. Pull down the Release Knob at the Skull Clamp
base and pull the Extension Assembly away from the Skull
Clamp base to widen the Skull Clamp.
2. Place the patient’s head in the desired position within the
Skull Clamp. Push the Extension Assembly into the Skull
Clamp Base to reduce the Skull Clamp width.