7.2 Automated Cleaning
Place the precleaned and dismantled products in an OP rack and start the automated cleaning
and disinfection procedure:
1. 2 min pre-washing with cold water
2. Drain
3. 5 min cleaning with mildly alkaline cleaner (like 0,5% neodisher MediClean, Dr. Weigert)
at 55°C/131°F
4. Drain
5. 3 min neutralizing with cold water.
6. Drain
7. 2 min intermediate rinsing with cold water
8. Drain
Thermal Disinfection has to be processed according national requirements and to the A0- value
according to ISO 15883.
Dryimmediatelyafternalrinse.Usethedryingcycleofthewasher/desinfectorand-ifrequired
-acleanlint-freeclothfordrying.Dryinternalareaswithlteredcompressedair,ifavailable.
Inspect each component for remaining debris; if any are present, repeat the cleaning procedure
using fresh detergent. Assemble all components as described in Section Mounting and Mainte-
nance. Inspect all devices prior to cleaning or storage to ensure instruments are suitable for use.
Store the device in a clean, dry, moisture free area at a temperature of 5°C/41°F to 40°C/104°F.
Important:
Please refer to chapters Mounting and Maintenance for dismantling information.
When dismantling the device for cleaning purposes and storage, make sure to
restore to the original state of the system.
Important:
Do not immerse radiolucent parts longer than 12 minutes in the cleaning solution
as it may cause swelling of the material.
Exclusion of liability:
The manufacturer shall not be liable for damages resulting from improper
cleaning and handling of the product and its parts.
Danger:
Never steam sterilize the device. If you autoclave the device, the heat will
damage the internal components and may damage the exterior nish.