DORO Retractor Arm:
Do not lubcricate link area. Lubricate all
movingpartsexcepttheexiblesegments
of the Retractor Arm.
7.3 Lubrication
Lubrication on all UNCOATED moving parts and threads of the ALUMINUM parts (exceptions
see below) should be done after every wash; failure to lubricate the Aluminum Headrest System
and Retractor System as recommended will result in a greatly reduced life of the equipment.
Please pay attention to all screws and threads. You can lubricate with any medical grade
lubricantformetalsandTeon.
Do NOT lubricate RADIOLUCENT parts and TEFLON coated parts.
Important:
The cylinder of the Base Unit Transitional Member, the Ball and the Socket
area of the Ball Pivot Adaptor and the exible segments of the retractor arms
should NEVER be lubricated.
DORO Base Unit:
The cylinder of the Base Unit Transitional
Member should NEVER be lubricated.
However, discrete lubrication of the ball
bearings on the cylinder of the Transitional
Member should be done, as long as the
lubricant does not contact the rest of the
cylinder. Do not lubricate the Connecting
Tube.
DORO Ball Pivot Adaptor:
The Ball and the Socket area of the Ball Pi-
vot Adaptor should NEVER be lubricated.
Warning:
Radiolucent parts and Teon coated parts should NEVER be lubricated.
Water-based lubricant on radiolucent parts may cause swelling and ANY
type of lubricant on radiolucent parts, especially on POM material will prevent
proper tightening of connections.
Warning:
Failure to properly lubricate as instructed may cause uncoated Aluminum
moving parts to dry out and be damaged.