Instruction manual
5.2
Mount the Swivel Adaptor to the Transitional Member of the Base Unit:
1. Position the teeth of the starburst connection of the
Transitional Member and the teeth of the lower starburst
connection of the Swivel Adaptor in a manner that they
engage.
2. Fully tighten the lower locking screw of the Swivel Adaptor
by turning it clockwise. Note: the teeth of the starburst con-
nections must be fully engaged and seated properly.
5.3
Mount the Skull Clamp to the Swivel Adaptor
1. Position the teeth of the upper starburst connection of the
Swivel Adaptor and the teeth of the starburst connection of
the Skull Clamp in a manner that they engage.
2. Position the Skull Clamp as desired.
3. Fully tighten the upper locking screw of the Swivel Adaptor
by turning it clockwise. Note: the teeth of the starburst
connections must be fully engaged and seated properly.
Danger:
Check all fastening screws and the stability of the complete DORO Headrest
System before and after each clinical use.
5.4
This system requires three DORO Skull Pins of the same versions to be used. Please refer
to the instruction manual of the appropriate DORO Skull Pins.
-13-