Product Manual

First it will inflate the mattress to 30mm Hg and then enter alternating mode.
If 30mm Hg cannot be reached within 30 minutes, the unit will stay in static mode and
keep inflating. If 30mm Hg still cannot be reached within another 15 minutes, the
alarm will beep and its LED will come on.
Press the Alarm Reset button to mute the alarm. The alarm LED will start blinking.
Press the Alarm Reset button again to have the audible alarm and alarm LED on.
Step 3 fThe alternating cycle time and pressure level are preset to the middle settings.
Select from the touch panel to adjust the cycle time and pressure level.
Press the Static button to set it in static mode, and the Static LED will come on. The
pump will enter static mode after it completes the last alternating cycle time. Press the
Static button again to restart the alternating function.
NOTE:
Press the static mode button from the touch panel to provide a firm surface that
makes it easier for the patient to transfer or reposition. The static mode can
support the patient by keeping the mattress from bottoming out when the
patient is in a sitting position.
Step 4 fDuring normal operation, the unit will monitor pressure. If the pressure is lower than a
set pressure, the pump will automatically inflate the mattress to the set level and then
stop. The alarm will beep and its LED will come on to alert if there is a low pressure
condition.
Press the Alarm Reset button to mute the alarm. The alarm LED will flash.
Press the Alarm Reset button again to have the audible alarm and alarm LED on.
Step 5 fPress the Auto Firm button to automatically inflate the mattress to the maximum level
(above 70 mmHg) for about 30 minutes. The pressure will return to a previously set
level after 30 minutes.
In case there is a bad connection of tubing or any air leakage from the mattress, this
will result in a low pressure condition, and it will activate the alarm.
NOTE:
For suitable pressure, please refer to page 9 for the hand check procedure.
CPR function
When there is an emergency to perform CPR on the patient, pull the CPR valves immediately
to release the air quickly from the mattress. The CPR valves are located at the front right-hand
side of the mattress.
Pressure set up
Users can adjust the pressure value of the air mattress to a desired firmness by themselves or
according to the suggestion from a health care professional.
NOTE:
It is recommended that the pressure-selector knob set to Firm or press Auto
Firm on the touch panel every each time the mattress is first inflated. Users can
then easily adjust the air mattress to a desired firmness according to the
patient’s weight.