Product Manual
sides, including all the connecting tubes and hoses. Let it sit for least 10 minutes. )
f If there is a base after you remove all the air cells, the base has to be sprayed down
with the disinfectant, inside and outside. Let it sit for the required incubation time and
wipe down with a cloth.
f Repeat the process with the tubing set: spray, incubate, and then wipe clean.
f The carrying bag should be turned inside out and completely wiped down using the
disinfectant solution.
f Allow it to thoroughly air dry. Once the inside is dry, turn it back; wipe down the outside
of the bag with disinfectant.
f Dry the mattress on a SUNLESS area after cleaning.
HANDLING AND STORAGE
f Lay the mattress out flat and upside down.
f Roll from the foot end towards the head end; the foot-end strap can then be stretched
around the rolled mattress to prevent unrolling.
f Do not fold, crease or stack the mattress.
MAINTENANCE
General
f Check the power cord and plug to see if there are abrasions or excessive wear.
f Check the mattress cover for signs of wear or damage. Ensure the mattress cover and
tubes are connected together correctly.
f Plug in the pump unit and check the airflow from the hose connection port. The airflow
should alternate between ports every half-cycle time.
f Check the air hoses to see if there are kinks or breaks. For replacement, please
contact your local agent or dealer.
f Make sure the mattress tube is well connected.
f Check the pump unit and make sure both power and power indicator are off when the
switch is turned off.
Low pressure
Examine if there is air leakage between the pump and the mattress connections or from the
air mattress tubes:
f Check connectors between the air mattress and pump. If there is any disconnection,
please reconnect it.
f Check the CPR Valves. Ensure their outlets are sealed.
f Check the air-connecting tubes. Ensure each single cell is not broken.
f Set the pressure-selector knob to Firm position. Keep the tubes fully inflated and
inspect for air leakage.
f Check if there is any air leakage from cells. Ensure no leakage occurs. If any leakage
occurs, please contact your local agent or dealer.