Product Manual
TROUBLESHOOTING
Problems Solutions
The pump doesn’t work
1. Check if the plug is inserted firmly into the outlet.
2. Turn on the power switch again.
f If the power LED is ON and the pump still doesn’t work, contact
your local dealer immediately.
f If the power LED is OFF, there may be a faulty outlet. Try to
connect the power cord to another outlet. If the power LED is still
OFF, contact a qualified electrician for main power check.
Incomplete inflation
(Low Pressure)
1. For a quick check, adjust the pressure to Firm position.
2. Check to see if the tubes connected to the pump are twisted or
there is any leakage occurring.
f Always keep the tubes straight.
f Change tubes if there is any leakage.
f Ensure the CPR valves are closed. Ensure every single cell is
not broken.
Slow air flow
A dirty filter may decrease the air flow.
Wash the filter with mild detergent to keep it clean.
f Check the air filter at the back of the pump unit at least once a
month.
SPECIFICATIONS
System
Item Specifications
Power Supply AC 120V/60Hz, 10W
Pressure Range f 30 ~ 60 mmHg 14029
f 10 ~ 70 mmHg 14030
Cycle Time f 10 min/60Hz 14029/14030
Dimensions 26.0 × 13.0 × 8.0 cm
10.2 x 5.1 x 3.1 in
Weight 1.5 kg
3.3 lbs
Environment Requirements f Temperature: Operation 50℉~95℉
Storage 59℉~122℉
Shipping 59℉~158℉
f Humidity: Operation 20%~80% non-condensing