Check and Go Disposable Operating Manual © 2005 Medical Depot, Inc Port Washington N.Y. 11050 visit us at www. drivemedical.com Check and Go Disposable Part# 18580 © 2005 Medical Depot, Inc Port Washington N.Y. 11050 www. drivemedical.
Effect of Interferent Gases and Vapors WARNING Volume % Dry Before use, all individuals who will be using this product must become thoroughly familiar with the information contained in this Operation Manual. Strict adherence to the operating instructions is necessary for safe and effective product performance. This product will perform only as designed and only if installed and operated in accordance with the manufacturer's operating instructions.
General Specifications Sensor Type: Drive galvanic cell w/Temperature Compensation (Non-Replaceable) Measurement Range: 0.0 - 99.9% oxygen (gas). Table of Contents Page Declaration of Conformity ………………………………. 1 Warning and Cautions ………………………………….. 2 Introduction ………………………………………………. 4 Features and Functions ……………………………. …. 5 Resolution/ Display: 0.1% - The three digit LCD indicates values between 0.0 - 99.9% oxygen.
Introduction The Check and Go Disposable is designed to monitor oxygen concentration in the patient-breathing environment. It is one of a full line of oxygen analyzers. The Check and Go Disposable utilizes the Drive oxygen sensor and is engineered for fast response, maximum reliability and stable performance. The Check and Go Disposable is designed primarily for spotchecking of oxygen levels delivered by medical oxygen delivery equipment and respiratory care systems.
Factors Influencing Calibration Features and Functions The primary factors influencing the Check and Go Disposable are temperature, pressure and humidity. Effects of Temperature The Check and Go Disposable will hold calibration and read correctly within ±3% when in thermal equilibrium within the operating temperature range. The device must be thermally stable when calibrated and allowed to thermally stabilize after experiencing temperature changes before readings are accurate.
LCD Display: A 3-digit display provides a direct readout of oxygen 3. concentration in the range of 0 - 99.9%. The display is blank when the Check and Go Disposable enters its Sleep (power off) mode. The Check and Go Disposable will automatically enter the Sleep mode after approximately 1.3 minutes from the last time the ON button was pressed. It is recommended that you conduct the calibration procedure again to get another reading.
Problem-Solving Calibrating the Check and Go Disposable Problem-Solving Table Before You Begin Condition Likely Cause Recommended Action LCD goes on momentarily and then returns to blank display (Sleep mode). • ON button was not actuated. If condition continues after making sure ON button is actuated, then analyzer is Low Battery disabled. Replace analyzer with a new Check and Go Disposable. Contact Drive Customer Service to reorder. While analyzer on, LCD shows a decimal point (".
4. 5. Attach an open-ended reservoir to the end of the "T" piece. Then start the calibration flow of oxygen at 2 liters per minute. Operating the Check and Go Disposable Six to 10 inches of corrugated tubing works well as a reservoir. A calibration oxygen flow to the Check and Go Disposable of 2 liters per minute is recommended to minimize the possibility of obtaining a "false" calibration value. To Check the Oxygen Concentration of a Sample Gas: 1.