Product Manual

520
Do not use if. . .
Do not use the SmartDose Gas Mini+ Conserver if it:
It is not working properly.
Has been dropped or damaged.
Has been submersed in water.
Call a qualified Health Care Dealer Technician for
examination and repair.
To prevent high concentrations of oxygen:
Keep equipment in well ventilated areas.
Do not carry equipment under a coat or any form of clothing.
Turn off oxygen supply by clos
ing the cylinder valve when
not in use.
Unintended oxygen leakage. If any unintended leakage of
oxygen is detected, do not attempt to use the product. Turn
the product off. If leakage persists, place the product out
doors, open doors and windows to ventilate the area, and
notify your medical equipment provider or service
representative of this condition.
Over pressure relief valve. The SmartDose
Mini+ Gas
Conserver is equipped with a pressure relief valve to ensure
the user’s safety. When activated, this safety feature may
make a slight hissing noise. If this noise occurs and persists,
turn the cylinder off and troubleshoot the problem.
Always supervise children. Children should always be
supervised around the SmartDose Mini+ Gas Conserver.
Failure to do so may result in damage to the uni
t or personal
injury.
Use in accordance with your prescription. For your safety,
the SmartDose Mini+ Gas Conserver must be used in accor
dance with the prescription determined by your physician.
Not for use in emergency transport vehicles. The SmartDose
Mini+ Gas Conserver is not intended to be used in
emergency transport vehicles (ambulances or helicopters),
or other similar areas where high levels
of Electromagnetic
Interference (EMI) may be expected.
WARNING
WARNING
WARNING
WARNING
WARNING
WARNING
WARNING
Transporting & Storing SmartDose Mini+ Gas Conserver
Preparing for Shipping
Prior to shipping the SmartDose Mini+ Gas Conserver, remove the
batteries and remove from oxygen cylinders.
Storage
Store the SmartDose Mini+ Gas Conserver in a cool, clean, dry area
when not in use. Avoid storing in direct sunlight.
Battery disposal
Dispose of all batteries in accordance with local, state and federal
reg
ulations.
Specifications
Weight (with batteries) 15.55 ounces
Dimensions 4.2 x 3 x 5
Power Supply 2 Standard AA” alkaline batteries
Operating Temperature 18°F to 10F (-8°C to 40°C)
Operating Pressure (cylinder) 500 3,000 psig (34 to 207 bar)
Operating Humidity 10 95 % RH, non-condensing
Storage Temperature - 20° to 50°C
Storage Humidity 0 95% RH, non-condensing
Continuous Flow Rate 2 LPM + 0.25 LPM
Volume /
Patient Setting 16 ml +20%,-10%
SmartDose 1 Volume 16 ml, incremental to standard volume
SmartDose 2 Volume 32 ml, incremental to standard volume