User manual

Tamarisk
®
320
Software ICD
29
NOTE
Note: the baud rate for the USB port is fixed at 921600.
Note: there is no acknowledge of this command, any subsequent commands
must be at the new baud rate.
The factory default setting is 57600 baud. There is a non-volatile parameter to permanently
change this. If a different baud rate is desired, there must be an initial command to change
the baud rate at 57600 baud. Or the non-volatile-parameter (#34) must be set to the desired
baud rate.
Note: the baud rate for the USB port is fixed at 921600.
Command Format:
Table 43Baud Rate Set Command Format
Byte Value Description
1 0x01 Start of message.
2 0xF1 Baud Rate Set Command ID.
3 0x02 Parameters length.
4, 5
Unsigned integer.
See table below.
Baud rate ID value.
6 checksum Frame check sequence.
Table 44Baud Rate ID Values
Baud Rate ID Baud Rate Baud Rate ID Baud Rate
0 230400 8 76800
1 115200 9 38400
2 57600 10 19200
3 28800 11 9600
4 14400 12 4800
5 7200 13 2400
6 3600 14 1200
7 1800 15 600
3.1.9 System Status Get 0xF2
Description: This command requests system status. The response contains information
about a variety of system functions.