36
Customers disabling automatic field calibrations must use command 0x25
Automatic Calibration Pending Activity Query and allow some calibrations to
maintain image and uniformity.
Command Format:
Table 61Automatic Calibration Activity Control Command Format
Byte Value Description
1 0x01 Start of message.
2 0x26 AutoCal Activity Control command ID.
3 0x02 Parameters length.
4, 5 0x0000 or 0x0001 Enable (0x0001) or Disable (0x0000)
6 checksum Frame check sequence.
Response Format:
Table 62Automatic Calibration Activity Control ACK Response Format
Byte Value Description
1 0x01 Start of message.
2 0x02 ACK Response
3 0x02 Parameters length.
4, 5 0x0026 AutoCal Activity Control Command ID.
6 checksum Frame check sequence.
3.2.7 Automatic Calibration Pending Activity Query 0x25
Description: This command queries the camera for pending Automatic Calibration
activities, and specifies the type of AutoCal that is pending. A suggested interval for
checking for pending calibration requests is every 5 minutes during normal operation and
every one (1) minute for the first ten (10) minutes immediately following power-up.
Command Format:
Table 63 Automatic Calibration Pending Activity Query Command Format
Byte Value Description
1 0x01 Start of message.
2 0x25 AutoCal Pending Activity Query command ID.
3 0x00 Parameters length.
4 checksum Frame check sequence.