User's Manual

Version: 1.0
least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power supply
lines
±1 kV for input/output
lines
±2 kV for power supply
lines
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
<<5% U
T
(>95% dip in U
T
)
for 0.5 cycle
0 % UT* for 0.5per at 0
°,45°,90°, 135°,
180°, 225°, 270°,
315°
0 % UT* for 1per
0% UT* for
5seconds(250/300per)
70 % U
T
* for 25/30per
<<5% U
T
(>95% dip in U
T
)
for 0.5 cycle
0 % UT* for 0.5per at 0
°,45°,90°, 135°,
180°, 225°, 270°,
315°
0 % UT* for 1per
0% UT* for
5seconds(250/300per)
70 % U
T
* for 25/30per
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the device requires continued
operation during power mains
interruptions, it is recommended
that the device be powered from an
uninterruptible power supply or a
battery.
Power frequency
(50Hz/60Hz) magnetic
field IEC 61000-4-8
30 A/m
30 A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.
NOTE U
T
is the a.c. mains voltage prior to application of the test level.
Guidance and manufacture’s declaration electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The customer or the user of the
device should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance
level
Electromagnetic environment - guidance
Conducted RF
IEC 61000-4-6
3 V
rms
150 kHz to 80 MHz
3 Vrms
6Vrms (emf)
in ISM and
amateur
radio bands
Portable and mobile RF communications equipment should be
used no closer to any part of the device, including cables, than
the recommended separation distance calculated from the
equation applicable to the frequency of the transmitter.
Recommended separation distance
80%, 1 kHz
Amp. Mod.