Remove the battery and replace it by a new fully charged battery. Connect the battery connector
from the new battery to the PCB connector.
NOTICE
Condensation in the device
Damage to device through formation of condensation if the temperature difference between
transportation or storage and the mounting location exceeds 20 °C (36 °F).
Before taking the device into operation let the device adapt for several hours in the new
environment.
NOTICE
Make sure that the wire running from the battery to the PCB is positioned on top of the battery,
i.e., on the opposite side of the PCB, to ensure correct operation of the Aloxy Pulse device.
Position the new battery on the support in the top shell and close the Aloxy Pulse again by repositioning
the bottom shell on top. Close the device by the 4 original screws, using a calibrated torque screwdriver
(with Torx T10 head) according to the table
7.4 To insure ingress protection IP69
WARNING
Loss of ingress protection IP69
Make sure that before assembly the tooth and groove is still resilient, clean and intact.
When assembling make sure that the gasket is seated correctly before tightening the bolds.
Only use LOW-HEAD SCREW WITH T-STER ISO ≈7380-1 STAINLESS STEEL A2 M3X8