Specifications

Chapter 3. Setting Up the DM5E
30 DM5E Operating Manual
3.4 Setting the Maximum and Minimum Alarms
When the thickness measurement value exceeds either the minimum or maximum alarm setting, the measurement value alternates between
solid and hollow font type displays, except in B-SCAN mode. When an alarm appears in B-SCAN mode, the measurement value flashes ON
and OFF in 1/2 second intervals.
To input a minimum or maximum alarm thickness:
1. Press to access the Configuration display (see Figure 9 on page 21).
2. Press or to select either MINIMUM ALARM or MAXIMUM ALARM.
3. When the desired parameter is highlighted, press to activate it.
4. Press or to change the selected position in alarm thickness and then press or to adjust the value of the selected
position. To switch the alarm setting to OFF from any other value, simultaneously press or .
5. Press after the alarm thickness has been set to the desired value.