Installation and Operating Instructions

2160100227L02 2001V008 27
CAUTION
Image plates are toxic
Image plates that are not packed in a
light protection cover can lead to poi-
soning when placed in the mouth or
swallowed.
i Only place image plates in the pa-
tient's mouth in a light protection cov-
er.
i Do not swallow the image plate or
parts of it.
i If the image plate or parts of it have
been swallowed, consult a specialist
doctor immediately and remove the
image plate.
i If the light protection cover has been
damaged in the patient's mouth, rinse
the mouth thoroughly with lots of wa-
ter. Do not swallow the water in the
process.
i Image plates must only be read using an im-
age plate scanner that is approved by Dürr
Dental.
i Image plates are flexible like X-ray film. How-
ever, the image plates should not be bent.
9 Correct use of image
plates
WARNING
Risk of cross contamination when
not using the light protection cover or
when using the light protection cover
more than once
i Do not use an image plate without a
light protection cover.
i Do not use the light protection cover
more than once (disposable item).
CAUTION
The image data on the image plate is
not permanent.
The image data is altered by light, natu-
ral X-ray radiation and scattered X-ray
radiation. This will lead to a reduction in
diagnostic information and clarity.
i Read the image data within 30 min-
utes of exposure.
i Never handle exposed image plates
without the light protection cover.
i Do not subject an exposed image
plate to X-ray radiation before or after
the scanning process.
i Do not X-ray during the scanning
process if the unit is in the same room
as the X-ray tube.
Usage
EN