Installation and Operating Instructions

30 2160100227L02 2001V008
Usage
i Completely slide the image plate into the light
protection cover. The black (inactive) side of
the image plate must be visible.
back
IQ
2
*+E24721301042100/$+01234567F*
back
IQ
2
*+E24721301042100/$+01234567F*
i Pull off the adhesive strip, fold down the flap
and close the light protection cover tightly by
pressing together firmly.
back
IQ
2
*+E24721301042100/$+01234567F*
back
IQ
2
*+E24721301042100/$+01234567F*
1
2
i The light protection cover must be disinfected
using a suitable disinfectant wipe immediately
before positioning it inside the patient's mouth
(see "11.2 Light protection cover").
back
IQ
2
*+E24721301042100/$+01234567F*
i Allow the light protection cover to fully dry.
10.2 X-ray
The procedure is described using a size
2 IQ image plate as an example.
Required accessories:
Image plate
Light protection cover the same size as the
image plate
WARNING
Risk of cross contamination when
not using the light protection cover or
when using the light protection cover
more than once
i Do not use an image plate without a
light protection cover.
i Do not use the light protection cover
more than once (disposable item).
WARNING
Danger due to re-use of products in-
tended for single use
The disposable item is damaged after
use and cannot be reused.
i Dispose of disposable items after use.
Preparing the X-ray
The image plate has been cleaned.
The image plate is not damaged.
The RFID tag is attached to the image plate.
If the RFID tag peels off, replace the image
plate.
i If using it for the first time or if it has been
stored for over a week: erase the image plate
(see "10.4 Erasing the image plate").
EN
2160100227L29_1910V008_US-EN_nur FCC-Zulassung_V2.indd 30 14.01.2020 11:04:13