Installation and Operating Instructions

32 2160100227L02 2001V008
Usage
Scanning the image plate
To avoid the mix up of X-ray images, only
scan the X-ray images from the selected
patient.
i Open the cover.
i Place the light protection cover with the image
plate centrally and straight onto the input unit.
The torn-off side of the light protection cover
faces down; the inactive (back) side of the im-
age plate faces the operator.
The image plate must not be pulled out of the
light protection cover before placement on the
input unit. There is the risk of image informa-
tion being erased by ambient light (see "9
Correct use of image plates").
back
IQ
2
*+E24721301042100/
$+01234567F*
back
IQ
2
*
+E24721301042100/$+01234567F
*
i Tear off the light protection cover.
back
IQ
2
*+E24721301042100/$+01234567F*
10.3 Scanning the image data
Starting the image plate scanner and soft-
ware
The reading-out process is described us-
ing the VistaSoft imaging software.
For further information on using the imag-
ing software, refer to the relevant manual.
i Press the on / off switch
to switch on the
device.
i Switch on the computer and monitor.
i Start VistaSoft.
i Select the patient.
i Select the corresponding image type in the
menu bar.
i Select the device.
i Set acquisition mode.
Recording starts directly.
Result:
The green status LED lights up. Scan the image
plate at this point (and not before).