Installation and Operating Instructions

2160100227L02 2001V008 35
Usage
i Soiling on both sides of the image plate
should be cleaned off with a soft, lint-free
wipe prior to every use.
i Remove resistant or dried on dirt with the im-
age plate cleaning wipe. When doing this, ob-
serve the instructions for use for the cleaning
wipe.
i Allow the image plate to completely dry before
using it.
11.4 Storage box with image plate
storage tray
Clean and disinfect the surface of the storage
box and the internal image plate storage tray in
the event of contamination or visible soiling.
Dürr Dental recommends the following disinfect-
ant for the storage box:
FD 366 sensitive
Dürr Dental recommends the following disinfect-
ants for the image plate storage tray:
FD 350 and FD 366 sensitive
i Clean the surface of the storage box and the
image plate storage tray with a soft, lint-free
cloth that has been dampened with cold tap
water.
i Disinfect the storage box using a disinfection
wipe. Alternatively, use disinfectant on a soft,
lint-free cloth.
i Disinfect the image plate storage tray using a
disinfection wipe.
Alternatively, the image plate storage tray
can also be treated in a thermal disinfector or
steam steriliser. Do not exceed a temperature
of 134°C when doing this.
NOTICE
Heat damages the cartridges
i Do not thermally disinfect or steam
sterilise the cartridges.
i Remove any soiling from the cartridge with a
soft, lint-free cloth that has been dampened
with cold tap water.
i Disinfect the cartridge using a disinfection
wipe. Alternatively, use disinfectant on a soft,
lint-free cloth. Comply with the operating in-
structions for the disinfectant when doing this.
i Dürr Dental recommends ID 213, ID 212 or
ID 212 forte instrument disinfectants for the
cartridges if an immersion disinfection system
is being used. Dürr Dental has subjected
these disinfectants to material compatibility
testing.
i Allow the cartridges to completely dry before
using them.
11.2 Light protection cover
The surface of the unit must be cleaned and
disinfected if it is contaminated or visibly soiled.
i Disinfect the light protection cover using a dis-
infectant before and after placement. Dürr
Dental recommends FD 333 forte wipes (viru-
cidal), FD 350 (limited virucidal activity) and FD
322 premium wipes (limited virucidal activity).
i Allow the light protection cover to completely
dry before using it.
11.3 Image plate
Cleaning and disinfection wipes are unsuitable
for cleaning image plates and may cause dam-
age to them.
Only use a cleaning agent that is compatible
with the materials:
Dürr Dental recommends the image plate clean-
ing wipe (see "3.4 Consumables"). Only this
product has been subjected to material compat-
ibility testing by Dürr Dental.
NOTICE
Heat or humidity will damage the im-
age plate.
i Do not steam sterilise the image plate.
i Do not immersion-disinfect the image
plate.
i Only use cleaning agents that are
compatible with the materials.