Instruction Manual

Page  25
Section 4 – Controls and Connections
Dukane Manual Part No. 403-583-00
Start-up Sequence
After all connections have been completed.
1. Push the Power Switch to ON (Figure 4-2).
The generator performs a self-diagnostics sequence.
2. Two Power-up screens appear briey - Figure 4-3,
and Figure 4-3A.
3. The next screen is an Operate screen ready for a
new weld to be done. The display shows:
The setup used for the last weld, and zeros for any
weld parameters.
See Figure 4-3B.
Figure 4-2 Power Switch
Push OFF
Push ON
Starting a Weld Cycle
1. If the generator is not powered, press its Power
Switch/Circuit Breaker to the ON position.
2. Select the setup you want to use, if appropriate.
3. The generator is ready to start a weld cycle when
the Ultrasound Activate/Cycle Start Input (Pin 12) is
activated.
Stopping the Weld Cycle
Normal Conditions
The cycle stops when the programmed welding cycle ends
if the generator is congured to weld by time or energy.
If it is congured to weld by automation, the cycle ends
when the Ultrasound Activate/Cycle Start Input (Pin 12)
is deactivated.
Emergency Conditions
Manual System
Push the Power Switch to OFF (See Figure 4-2.) to
stop the ultrasound signal. This may be done under any
conditions.
Automated System
Customer-supplied external controls provide the means
to stop the cycle for an automated system.
An auxiliary cable connects these external controls to the
iQ generator at the INPUTS/OUTPUTS connector.
Figure 4-3B Operate Screen, After Power-up
Weld by Time
Weld Time 0.000 S
Weld Energy 0 J
# 1
DUKANE
Setup #1
xxxxxxxxx
xxxxxx
http://www.dukane.com
iQ AL Series xxkHz
Figure 4-3 Power-up Screen 1
Software
version
Current
Setup
Website
Address
MB FPGA
version
Figure 4-3A Power-up Screen 2
U.S. Patent #7475801
Trigger by Power
available