Power Supply User Manual

Section 7 – Troubleshooting
Page  51
Dukane Manual Part No. 403-583-00
No Ultrasonic Output
Probe
Make sure that the probe cable is connected to the generator
connector (PROBE) and secured to the rear panel. Also,
make sure the probe stack is properly assembled.
System Power Output Level
Overload
When an overload occurs, it will automatically reset
when the next ultrasound activation signal begins. If the
condition persists:
Turn the generator OFF and:
1. Check the system. Change the probe to one that is
known to be good.
2. Turn the generator ON, and see if the fault condition
has been corrected.
Overtemperature
When the system overheats, and the generators internal
temperature exceeds 85°C (185° F) an overtemperature
fault condition will trip.
When the system cools, the system automatically resets
the overtemperature fault.
Generator Fault Does Not Reset
When a system fault condition does not automatically
reset, the generator needs servicing.
System Power Diagnostic
Procedures
The only fault indications available with these probe
systems are the ones shown on the LCD display.
NOTE
The LCD screen displays a variety of
pop-up status changes as they oc-
cur. Check Table 7-I - Pop-up Fault
Status Screens - Page 53.
NOTE
When Latching Faults is enabled,
ENTER must be pressed to clear
a fault.