Manual

Model Range Max. Press. Model Range Max. Press.
616-00 0-1 in. w.c. 2 psig 616-8 0-10 psid 29 psig
616-0 0-2 in. w.c. 2 psig 616-9 0-20 psid 58 psig
616-1 0-3 in. w.c. 2 psig 616-10 0-30 psid 58 psig
616-2 0-6 in. w.c. 5 psig 616-11 0-50 psid 150 psig
616-3 0-10 in. w.c. 5 psig 616-12 0-100 psid 150 psig
616-4 0-20 in. w.c. 11 psig 616-3B 1.5-0-1.5 in. w.c. 2 psig
616-5 0-40 in. w.c. 11 psig 616-6B 3-0-3 in. w.c. 5 psig
616-6 0-100 in. w.c. 29 psig 616-10B 5-0-5 in. w.c. 5 psig
616-7 0-200 in. w.c. 29 psig 616-20B 10-0-10 in. w.c. 11 psig
DWYER INSTRUMENTS, INC.
Phone: 219/879-8000 www.dwyer-inst.com
P.O. BOX 373 • MICHIGAN CITY, INDIANA 46361, U.S.A. Fax: 219/872-9057 e-mail: lit@dwyer-inst.com
The Series 616 Differential Pressure Transmitter senses the pressure
of air and compatible gases and sends a standard 4-20 mA output
signal. A wide range of models are available factory calibrated to specific
ranges as listed in the chart below. The span and zero controls are for
use when checking calibration. They are not intended for re-ranging to a
significantly different span. Versatile circuit design enables operation in
2, 3 or 4-wire current loops.
For applications requiring direct pressure readings or percent of full span
output, the optional Model A-701 Digital Readout makes an ideal
companion device. It provides a bright red 0.6˝ high, 3-1/2 digit LED
display while supplying power to the Series 616 transmitter. For
additional information on these and other transmitters, see the Dwyer
Instruments, Inc. Full Line catalog.
Series 616 Differential Pressure Transmitter
Bulletin E-43
Specifications – Installation and Operating Instructions
5/16
[7.950]
7/8
[22.23]
3-21/64
[84.53]
2-13/32
[61.12]
1/2
[12.70]
1/2
[12.70]
1-1/2
[38.10]
2-1/4
[57.15]
ø5/32 [3.97]
MOUNTING HOLE
TYP 2 PLACES
HI LO
SPECIFICATIONS
Service: Air and non-combustible, compatible gases.
Wetted Materials: Consult factory.
Accuracy: 616: ±0.25% F.S.; 616C: ±1.0% F.S.
Stability: ± 1% F.S./yr.
Temperature Limits: 20 to 120°F (-6.67 to 48.9°C).
Pressure Limits: See Chart.
Thermal Effect: 616: ±0.055% F.S./°F (0.099% F.S./°C); 616C:
±0.070% F.S./°F (0.125% F.S./°C).
Power Requirements: 10-35 VDC (2, 3 or 4 wire); 16-26 VAC (4 wire).
Output Signal: 4 to 20 mA.
Zero and Span Adjustments: Potentiometers for zero and span.
Loop Resistance: DC: 0-1250 ohms maximum; AC: 0-1200 ohms
maximum.
Current Consumption: DC: 38 mA maximum; AC: 76 mA maximum.
Electrical Connections: Screw-type terminal block.
Process Connections: Barbed, dual size to fit 1/8˝ and 3/16˝
(3.12 mm and 4.76 mm) I.D. rubber or vinyl tubing.
Mounting Orientation: Vertical, consult factory for other position
orientations.
Weight: 1.8 oz. (51 grams).
Series 616 Transmitter Models and Ranges*
*All models available with 0.25% F.S. Accuracy.
Models available with 1.0% F.S. Accuracy include 616-1 through 616-20B.
E-43:E-43 5/24/10 4:10 PM Page 1

Summary of content (4 pages)