OPERATION MANUAL ANEMO-PSYCHROMETER & BTU-PSYCHROMETER Model:8903/8904/8906 Model: 8902 8911 8912
INTRODUCTION Thank you very much for purchasing this Anemo-Psychrometer. This unique meter design with 7 H.V.A.C &R must parameters in 1. The meter designed as battery operated for Humidity ,Air temp., Dew Point, Wet Bulb, Air Velocity, Air Volume & BTU (8912). The sensor is built in the remote fan and is specially protected by tunable cap. While in operation, please turn on the cap to get accurate temperature and humidity reading. The psychrometer is a micro processorbased design.
MATERIAL SUPPLIED This package contains: The meter x 1 Remote vane x1 Battery x 4 (AAA size ) Operation manual x1 Hard carry case x1 Optional accessory: AC to DC 9V adaptor RS232 cable & CD REMINDER Important: 1. Rotate the cover before measurement to ensure the measured data is correct. 2. The fan and meter are sold as a kit and . are well calibrated. When you need to replace the fan, please see page 16 for the details of fan replacement.
KEY PAD 1.Power : -Turn on the meter with auto-sleep mode. -Turn off the meter at any mode. -When the meter is off, pressing more than two sec. to enter units selection. 2.MODE: -Pressing to select different modes. Temp. DP WB RH Velocity. -Pressing longer to select VOL and CAP. 3.ENTER: - To confirm the setting & calibration. 4.Mx/Mn/Up : -Pressing to view MAX/MIN/AVG value. -Press to select the value of each digit in cycle. 5.
6.HOLD/Down : - In basic modes pressing this key to hold the current reading, then pressing this key again to unlock the holding. - Pressing to select the setting digit. 7.Power + : - When the meter is off, pressing more than two sec. to enter non-sleep mode. 8.Power + + : - Pressing more than two seconds to enter RH calibration mode. 9.ENTER+ : - Pressing to turn on/off the backlight. POWER ON/OFF & UNIT SELECT Press to turn on the meter with auto-sleep mode. Press again to turn off the meter at any mode.
Basic Measure 8902 8911 8912 8911 8912 Air temp. Humidity Dew Point Wet Bulb Air Velocity Vel with average Extend Measure 8902 Air Volume Sum of Volume Capacity Sum of Capacity When power on, the air temperature is the default value to show on the LCD. To review the other basic parameters, just press key then each parameter will be display in turn. While in basic mode, press for over 2 seconds to enter extend measurement mode. For 8912, also press key to switch between volume and capacity.
MAX, MIN & AVG. From the meter is powered on, air temp., dew-point, WBT, humidity & velocity will be temporarily saved in meter, so, to press could view the average value minimum value and maximun value in turns. Press again to back to normal mode. The data will be re-counted once the meter is powered off and turn on again. When in Max./Min./Avg. mode, pressing key to display the parameters you need. (Fig. B, Fig. C) DP MAX MIN F %RH (Fig. C) (Fig.
MIN AVG Ft/m HOLD (Fig.D) Ft/m HOLD (Fig.E) AIR VOLUME: OUTLET SIZE While in normal mode, long pressing key to enter air volume mode. To measure the volume, the outlet size need to be input first. There are three choice to enter the size: - Length & Width: This is the default choice. Press key to select the setting digit first, then press key to select the appropriate value ( 0~9). When select the value, the value will show from 0 to 9 in cycle.
Press key to select the setting digit first, then press key to select the appropriate value ( 0~9) . After selection, short press again to finish the size setting and enter next step. (Fig. H) VOL VOL cm cm L W (Fig. F) VOL VOL cm 2 cm D AREA (Fig. G) (Fig. H) Note: Refer to page 13 for the range of each size input mode. AIR VOLUME: MEASUREMENT After short pressing to finish and leave size setting, there are 20 second waiting time for you to put the vane to the grill.
After starting measuring the volume, the meter will automatically count the volume for 60 seconds to give a average value. (Fig. J) During this 60 second time, the vane should move along the whole outlet to cover each area so the measured data could be more accuracy. The count down number displayed on the top- left corner as a reminder and meter also beeps when 60 sec is up. After time is up, 60 seconds average volume will display on the LCD as Fig K. AVG VOL VOL 3 M/m 3 M/m START (Fig. K) (Fig.
There are three choice to enter the size: - Length & Width: This is the default choice. Press key to select the setting digit first, then press key to select the appropriate value ( 0~9). When select the value, the value will show from 0 to 9 in cycle. After entering length, short press to enter width setting. Repeat the procedure of length setting, then short press again to finish the size setting and enter next step. (FIg.
CAP cm CAP 2 cm D AREA AREA (Fig. N) (Fig. O) CAPACITY: MEASUREMENT After short pressing to finish and leave size setting, there are 90 or 20 second (note 1) waiting time for you to put the vane to the grill. During above waiting time, the count down number displayed on top-left corner and meter also beeps when time is up.(Fig. P) If no need to wait for 90 or 20 seconds, just pressing key to start measuring the capacity right away.
AVG CAP Kw CAP Kw START (Fig. R) (Fig. Q) If there are more than one outlet, you could press key again to repeat the procedure from outlet size input. You could measure outlet as many as you need. After measuring more than one outlet volume, you could press key to review the sum of all outlets. Pressing key to review the last outlet capacity.
AC TO DC ADAPTOR The compatible AC to DC adaptor is 9V at least 200mA. AVAILABLE SIZE RANGE Please refer to below input size range. Please be reminded that in metric L/W/D , the value after decal is not selectable. AREA L/W/D cm 000.0~999.0 00000~99999 inch 000.0~999.9 00000~99999 RS232 SOFTWARE Software We have compatible software and cable which is design for the meter, please contact with your supplier for the details. Interface of RS232 A.9600 bps, 8 data bits, no parity. B.Format: Tx.
Where: The 1 st value is velocity, the 2 nd value is Air temp., the 3 rd value is Humidity, the 4 th value is Dew Point, the 5th value is Wet bulb, the 6 th value is AIR Volume, the 7th value is Capacity. The x here means one of {0|1|2| |9|-} C.Format for error value: E01 Probe didn't connected; E02 Under flow; E03 Over flow. See more in page The unit for error code is Nul. EX: V010.5MPS:TE02Nul:H66.7%: dE04Nu:wE04Nul:v00020.5CMM. TROUBLESHOOTING 1.
5.E 3 a)The value is overflow. 4.E4 a)The original data that is relative to this value error. 5.E5 a)Out of meter display range. 6.E6 a)the value is not calculated completely. 7.E 11 a)Humidity Calibration error. SPECIFICATION O O O TEMPERATURE : -20 C to 60 C (-4 to 140 F) O O Accuracy.:+0.6 C(+1 F) , O O Resolution :0.1 C (0.1 F) RELATIVE HUMIDITY : 0 to 100% RH Accuracy :+3% at 10 to 90%RH (cal. temp.) +5% at other range Resolution : 0.1% O O DEW POINT : -68 to 70 C (-90 to 158 F) O Resolution:0.
ADDITIONAL FEATURE : BTU CAPACITY : 0 to 99999 BTU/H (KW) (8912) Resolution : 0.1 ( 0 - 9999.9 ) or 1 (10000 - 99999) Dimension : (HxWxD) Vane:170 x 77 x 40 mm3; Meter:175 x 70 x 33 mm Power:4 x 1.5V AAA battery or 9V >200mA adaptor. FAN REPLACEMENT If users have more than one set anemometers or users change a new fan for his old meter, user should input the RH% calibration data which enclosed with each fan.
Important Note: While setting the slope, only input 5 digits after decimal point, so user need to round off the 6th digit. User should also input two digits before decimal. Example: a. If the S value is 8.239678, user should input 08 23968 b. If the S value is 11.23968, user should input 11 23968 c. If the S value is 0.2396838, user should input 00 23968 (Fig. S) (Fig. S) Intercept (value Z) 1. While press key save the S setting, the meter will enter intercept input automatically.
User should also input four digits before decimal. All the Z is a negative value but LCD will not appear the negative icon. Example: a. If the Z value is -928.8683, user should input 09 28868 b. If the Z value is -1928.868, user should input 19 28868 c. If the Z value is -89.68482, user should input 00 89685 (Fig. T) (Fig. T) RETURN AUTHORIZATION Authorization must be obtained from the supplier before returning items for any reason.
Accuracy, the Zenith of Measuring / Testing Instruments ! Hygrometer/Psychrometer Thermometer Anemometer Sound Level Meter Air Flow meter Infrared Thermometer K type Thermometer K.J.T. type Thermometer K.J.T.R.S.E. type Thermometer pH Meter Conductivity Meter T.D.S. Meter D.O. Meter Saccharimeter Manometer Tacho Meter Lux / Light Meter Moisture Meter Data logger Temp./RH transmitter Wireless Transmitter .......... More products available ! 2010.