Manual

Data Logger Manual
Page 3
Section 1: Device Safety
CAUTION: DEVICE MUST BE USED ONLY IN A MANNER CONSISTENT
WITH THIS MANUAL.
NOTICE: WHEN 230VAC SUPPLY USED, DWYER SPECIFIES USE OF
AC SUPPLY PART # T35-9-100R-3 MANUFACTURED BY ENG
ELECTRIC. THIS SUPPLY IS AVAILABLE FROM DWYER.
1.1 Model information
• Model: DLI-R
Description: Single channel RTD logger with LCD display
Manufactured in the USA
1.2 Specifi cations
External Probe
Range: -328 to 1562ºF (-200 to +850ºC)
Resolution: 0.01ºF (0.01ºC)
Calibrated accuracy (case between 59 to 95ºF (15 to 35ºC))
±0.05ºF (±0.05ºC) Probe between -328 to 500ºF (-200 to 260ºC)
±0.1ºF (±0.1ºC) Probe between 500 to 1562ºF (260 to 850ºC)
Accuracy de-rating:
0.05ºF (0.05ºC) per ºF (ºC) outside 59 to 95ºF (15 to 35ºC ) case
temperature, probe between -328 to 500 (-200 to 260ºC)
0.1ºF (0.1ºC) per ºF (ºC) outside 59 to 95ºF (15 to 35ºC) case
temperature probe between 500 to 1562ºF (260 to 850ºC)
Dot-Matrix LCD
Dimensions: 2.5” x 1.375” (63mm x 35mm)
Text: Confi gurable channel text size
Indicators: Power, status, memory
Backlight: Confi gurable w/auto shut-off and contrast adjustment
Start/Stop Time:
Software programmable start time and date, up to six months in
advance; programmable stop time
S
tart
/
Sto
p
Time
:
So
f
tware
p
ro
g
ramma
bl
e start time an
d
d
ate, u
p
to six mont
h
s in
a
d
vance;
p
ro
g
ramma
bl
e sto
p
tim