User Manual
Particulate Monitoring Systems  Installation & Operating Manual 
Document No. 210-1003-M  Page 25  ©2007 
Once the control unit and coaxial cable zero have been checked, proceed to the Sensor Zero Check. 
To perform the sensor zero check the process flow must be stopped or a sensor test pipe (available 
from Factory) or length of metal pipe will be needed (4”-6” diameter pipe or larger). The pipe should 
be at least 3 in (8 cm) longer than the probe itself and must be grounded. The length of pipe will serve 
as an electrical shield for the probe while it is out of the process. 
SENSOR ZERO CHECK 
1.  Do not remove the sensor from a running process if it will in any way 
compromise personnel, plant safety or hazardous area safety procedures. 
2.  Disconnect power to the control unit. 
3.  Remove the sensor from the process and insert it into the grounded metal 
test pipe. For hazardous area sensors do not disconnect the external sensor 
earth ground strap. 
4.  Re-apply power to the control unit and allow the reading to stabilize for 1-2 
minutes. 
5.  Read the display. It should be below the control units specified minimum 
detection level. If the sensor passes this check there are no false signals 
from the sensor. 
PASS: 
1.  Disconnect power to the control unit. 
2.  Remove the sensor from the grounded test pipe and re-insert into the 
process. For hazardous area sensors do not disconnect the external sensor 
ground strap. 
FAIL: 
1.  Contact the factory for further assistance. 
When performing a zero check, keep in mind that it may be acceptable to consider a small false signal 
negligible. For example if the baseline readings are 100pA and a system zero offset of 1pA was 
found, this is only a 1% affect on the normal readings. If using the device for basic flow/no flow 
detection or basic emissions detection, this would not be significant. 










