Instruction Manual

Appendix: Meter Factory Default Setting
MAINTENANCE/REPAIR
Upon final installation of the Series EC2-10, no routine maintenance is required.
The Series EC2-10 is not field serviceable and should be returned if repair is
needed. Field repair should not be attempted and may void warranty.
W
ARRANTY/RETURN
R
efer to “Terms and Conditions of Sales” in our catalog and on our website. Contact
customer service to receive a Return Goods Authorization number before shipping
the product back for repair. Be sure to include a brief description of the problem
plus any additional application notes.
DWYER INSTRUMENTS, INC.
Phone: 219/879-8000 www.dwyer-inst.com
P.O. BOX 373 MICHIGAN CITY, INDIANA 46360, U.S.A. Fax: 219/872-9057 e-mail: info@dwyer-inst.com
©Copyright 2013 Dwyer Instruments, Inc. Printed in U.S.A. 6/13 FR# R6-444016-00 Rev. 2
Symptom
Power on but no display
Display disappeared
A
ir bubbles adhere on
e
lectrode
Action
1. Make sure the time of pressing the power key is
more than .3 seconds.
2
. Check that the batteries are in place and make good
c
ontact, minding polarities.
3
. Replace new batteries.
4
. Move the battery away for one minute and then put
back in.
Check whether the low battery icon was on before the
display went off. If yes, replace with new batteries.
1
. Stir the electrode completely and dip it into the
s
olution at an oblique angle. Dipping the electrode in
v
ertically creates many air bubbles.
2
. After soaking the electrode in solution for 15 to 30
minutes, inspect the electrode carefully to make sure
no air bubbles adhere. If they still exist, tap the bottom
of the container gently and stir the electrode to remove
t
he air bubbles.
3
. If the above method is not working, remove the
e
lectrode out of the solution and blow at the electrode
to remove the air bubbles.
Error Code Meaning
Parameter: Conductivity
---
E03
E04
Meter is in manual ranging 1;
however, the conductivity
measured value is higher than
1999uS.
Conductivity value is over the
range limit (19.99mS) or meter is
damaged.
The original temp. error results in
this error.
Press “UP” key more than 2
seconds to change the mode to
manual ranging 2 or auto
ranging.
Put the meter in standard buffer
(the buffer concentration must be
lower than range limit). If E03 still
appears, send back for repair.
Refer to below error code of
temp. After solving the error of
temp, E04 of conductivity will
disappear.
How to Handle?
Parameter: Temperature
E
01
E02
E
03
T
emperature circuit is damaged.
T
emperature value is lower than
r
ange limit (0°C) or temperature
c
ircuit is damaged.
Temperature value is higher
than range limit (50°C) or
t
emperature circuit is damaged.
S
end back for repair.
P
ut the meter in room
temperature for 5 minutes to
make the readying back to
normal. If E02 still appears,
s
end back for repair.
P
ut the meter in room
t
emperature for 5 minutes. If
E
02 still appears, send back for
repair.
T
ype
P
1.1
P
1.2
P
1.3
P3.1
P4.1
P
4.2
P
arameters
S
elect °C/°F
N
or. Temp.
T
emp. Coefficient
Factory Default
Viewing Previous
C
alibration Data
D
efault
°
C
2
5°C
2
.1%/°C
NO
---
-
--
R
emark
T
emperature Unit
S
electable: 25 or 20°C
A
djustable from 0.4 to 10%
Retain Your Current Settings
No Calibration Data for 1st
Range
N
o Calibration Data for 2nd
R
ange