LCR10 IOM  Page 16 of 20  April, 2003
Chapter 4 Calibration
4-1 Pen Arm Calibration
Pen calibration is a two point calibration. To check and/or adjust the
recording pen(s) calibration to the inner and outer graduations of the 
chart,
Perform the following.
1.  Remove power to instrument and remove the battery.
2.  With all power removed, press and hold the CHANGE CHART key 
(3).
  While holding the CHANGE CHART key(3), re-apply power. Release 
key upon applying power, either AC or battery, which ever is 
convenient.
3.  The pen will move off scale, The LED will fl ash rapidly. The Green 
LED will then go to steady on.
4.  The red pen will then move to the 100% ring. Adjust pen using the 
LEFT (1) and RIGHT (2) keys to position pen on the 100% ring, if 
necessary, Then Wait.
5.  The Red pen will then move to the 0% ring. Adjust the pen position 
using the LEFT (1) and RIGHT (2) keys, if necessary, then wait.
6.  The Red pen will then move full upscale.
8.  Red pen calibration is now complete. If unit has alarms enabled pens 
will now go to current alarm settings. See Section 3.2. Then to actual 
recording value.
  If no alarms are enabled pen will then go directly to actual recording 
values and begin normal operation.
Each time the chart paper or fi ber tip pen cartridge is changed, you 
should make sure that each pen stops at the inner and outer most 
temperature graduations of the chart paper. Otherwise, this pen offset 
can cause the unit to record an incorrect value on the chart.