Internal Source Calibration (ICAL) 
The meter is factory calibrated prior to shipment 
to read in milliamps and volts depending on the 
input selection. The calibration equipment is 
certified to NIST standards. 
The use of calibrated signal sources is necessary to calibrate the 
internal source of the meter. The meter’s internal source is what allows 
the user to scale the meter without applying a signal. 
Check calibration of the meter at least every 12 months. Each input 
must be recalibrated separately. 
Notes: 
1.  If meter is in operation and it is intended to 
accept only one input type (e.g. 4-20 mA), 
recalibration of other input is not necessary. 
2.  Allow the meter to warm up for at least 
15 minutes before performing the internal 
source calibration procedure. 
The 
Internal calibration menu is part of the Advanced Features menu. 
1.  Press and hold the Menu button for three seconds to access the 
advanced features of the meter. 
2.  Press the Up
 arrow button to scroll to the Internal calibration menu 
(
ICAL) and press Enter. 
3.  The meter displays either current calibration (
C CAL) or voltage 
calibration (
v CAL), according to the input setup. Press Enter to 
start the calibration process. 
Example of Internal Calibration for current input: 
4.  The meter displays low input current message (C lo). Apply the 
low input signal and press Enter. The display flashes for a moment 
while the meter is accepting the low input signal. 
5.  After the display stops flashing, a number is displayed with the 
leftmost digit brighter than the rest. The bright digit is the active digit 
that can be changed by pressing the Up
 arrow button. Press the 
Right
 arrow button to move to the next digit. 
6.  Set the display value to correspond to the input signal being 
calibrated, typically 4.000 mA. 