User guide
Multi Process Meter   Instruction Manual 
82
Meter Copy or Cloning Instructions 
Caution! 
Do not connect the two meters to 
the same signal source while 
cloning. Internal calibration may 
be affected. 
1.  Connect two meters using a PMA-02 meter copy cable. 
Warning! 
Using standard CAT5 or other cable will cause 
damage to both meters. Use PMA-02 meter copy 
cable only. 
2.  Do not connect the two meters to the same signal source. 
3.  Power up both meters. Leave Clone meter in Run Mode. 
4. Enter the Advanced Features menu of the Master meter; see 
Advanced Features Menu on page 67. 
5.  Scroll to the Copy function using the Up arrow button then press 
Enter. 
6.  The meter displays the message Send. Press Enter, the display 
flashes while sending data. The message done is displayed when 
copying is completed. 
7.  The Clone meter displays the message CPY rC while being 
programmed then the message done when copying is completed. 
The meter initializes and returns to Run Mode using the same 
settings as the Master. 
8.  If meter to be cloned does not respond to the data being sent, refer 
to Copy Function Requirements above. 
!