Owner's manual
O
ption 2: Using Conversion Factors
T
DS values are related to conductivity. You can calibrate the meter by using
c
onductivity standards as described above and then program the meter with a
g
iven conversion factor. Please refer to below steps.
1. Perform the conductivity calibration procedure.
2. Select the correct conductivity-to-TDS conversion factor for other solutions using
the formula: Factor = Actual TDS = Actual Conductivity @ 25°C.
3
. When the meter is in measurement mode, press “set button” for more than 2
s
econds. Press “up” to get to P2.0. In P2.0 press “set button picture”
m
omentarily to enter P2.1. The factor will flash on the LCD. Press “up” or “down”
t
o change factor from 0.40 to 1.00. Press “set button picture” momentarily to
confirm the value and the meter will return to P2.0.
Appendix: Meter Factory Default Setting
MAINTENANCE/REPAIR
Upon final installation of the Series TDS2-10, no routine maintenance is required.
The Series TDS2-10 is not field serviceable and should be returned if repair is
needed. Field repair should not be attempted and may void warranty.
WARRANTY/RETURN
Refer to “Terms and Conditions of Sales” in our catalog and on our website. Contact
customer service to receive a Return Goods Authorization number before shipping
the product back for repair. Be sure to include a brief description of the problem
plus any additional application notes.
DWYER INSTRUMENTS, INC.
Phone: 219/879-8000 www.dwyer-inst.com
P.O. BOX 373 • MICHIGAN CITY, INDIANA 46360, U.S.A. Fax: 219/872-9057 e-mail: info@dwyer-inst.com
©Copyright 2013 Dwyer Instruments, Inc. Printed in U.S.A. 6/13 FR# R6-444017-00 Rev. 2
Symptom
P
ower on but no display
D
isplay disappeared
A
ir bubbles adhere on
electrode
Action
1
. Make sure the time of pressing the power key is
m
ore than .3 seconds.
2
. Check that the batteries are in place and make good
contact, minding polarities.
3. Replace new batteries.
4. Move the battery away for one minute and then put
back in.
C
heck whether the low battery icon was on before the
d
isplay went off. If yes, replace with new batteries.
1
. Stir the electrode completely and dip it into the
solution at an oblique angle. Dipping the electrode in
vertically creates many air bubbles.
2. After soaking the electrode in solution for 15 to 30
minutes, inspect the electrode carefully to make sure
no air bubbles adhere. If they still exist, tap the bottom
of the container gently and stir the electrode to remove
the air bubbles.
3. If the above method is not working, remove the
electrode out of the solution and blow at the electrode
to remove the air bubbles.
Error Code Meaning
Parameter: Conductivity
---
E
03
E04
M
eter is in manual ranging 1;
h
owever, the conductivity
m
easured value is higher than
1
999uS.
Conductivity value is over the
range limit (19.99mS) or meter is
damaged.
T
he original temp. error results in
t
his error.
P
ress “UP” key more than 2
s
econds to change the mode to
m
anual ranging 2 or auto
r
anging.
Put the meter in standard buffer
(the buffer concentration must be
lower than range limit). If E03 still
a
ppears, send back for repair.
R
efer to below error code of
t
emp. After solving the error of
t
emp, E04 of conductivity will
disappear.
H
ow to Handle?
P
arameter: TDS
-
--
E04
M
eter is in manual ranging 1;
h
owever, the TDS measured
value is higher than 1999*TDS
factor ppm.
The original temp. or
c
onductivity error results in this
e
rror.
P
ress “UP” key more than 2
s
econds to change the mode to
manual ranging 2 or auto
ranging.
Refer to below error code of
t
emp. & conductivity. After
s
olving the error of temp, &
c
onductivity, E04 of TDS will
disappear.
Parameter: Temperature
E
01
E02
E03
T
emperature circuit is damaged.
T
emperature value is lower than
range limit (0°C) or temperature
circuit is damaged.
T
emperature value is higher
than range limit (50°C) or
temperature circuit is damaged.
Send back for repair.
Put the meter in room
temperature for 5 minutes to
make the readying back to
n
ormal. If E02 still appears,
s
end back for repair.
P
ut the meter in room
temperature for 5 minutes. If
E02 still appears, send back for
repair.
Type
P1.1
P1.2
P1.3
P2.1
P3.1
P4.1
P4.2
Parameters
Select °C/°F
Nor. Temp.
Temp. Coefficient
TDS Factor
Factory Default
Viewing Previous
Calibration Data
Default
°C
25°C
2.1%/°C
0.50
NO
---
---
Remark
Temperature Unit
Selectable: 25 or 20°C
Adjustable from 0.4 to 10%
Adjustable from 0.40 to 1.00
(only for 8361)
Retain Your Current Settings
No Calibration Data for 1st
Range
No Calibration Data for 2nd
Range
If the standard buffer is over the measuring limit or less than
10% of the measuring limit, the displayed value will equal the
range limit or 10% of the range limit.
W
hen switching the meter from measurement to calibration mode, the meter will
d
isplay the factory default value. If the meter was previously calibrated, the display
may seem to jump to the factory default value when entering calibration.
NOTICE