User Manual
20of30 MN890570 RevA
INSTALLATION StepNine
TheSeriesULSshouldalwaysbemountedperpendiculartotheliquidsurfaceandinstalledusingtheprovided
FKMmountinggasket.Makesurethatthefittingandtransmitterthreadsarenotdamagedorworn.Always
hand‐tightenthetransmitterwithinthefitting.Performaninstalledleaktestundernormalprocessconditions
priortosystemstartup.Note:ThepreferredmountingfittingfortheSeriesULSS&ULSMistheULS‐ACC‐121
(2”threadx1”thread)reducerbushing.
MountingGuide
1. Donotmountatanangle
2. Liquidshouldneverenterthedeadband
3. SideWall:
a. Mountat
least2”fromthesidewall
b. ForSeriesULSL‐mountatleast3”fromthesidewall
4. Donotmountwhereobstacleswillintrudeonsensor’sbeamwidth
a. SeeSpecifications,StepTwo
5. Donotmountinavacuum
6. Avoidmountinginthecenterofadome
toptank.
7. Inconebottomtank,positionthesensoroverthedeepestpartofthetank.
Donotinstallat
anglerelativetothe
liquid.
Donotinstallwithin
2”oftanksidewall.
(3”forSeriesULSL)
Donotinstallwith
objectsinthebeam.
Donotinstallin
applicationswith
vacuum.
Installation in existing f ittings:If the existing fitting is larger than the threads of the Series ULS, select a
reducerbushingsuchastheULS‐ACC‐121(2”threadx1”thread)orULS‐ACC‐131(3”threadx2”thread).
MetalTanks(SeriesULSS&ULSM):DwyerInstruments,Incultrasonictransmittershavebeenoptimizedfor
useinnon‐metallicfittings.
1. Forbestperformance,avoidtheuseofmetallicfittings.
a. Useaplastic2”x1”reducerbushing,suchastheULS‐ACC‐121oraplastic1”flange,suchasthe
ULS‐ACC‐312formetallictanks.
2.
While installations directly into a 1” metal fitting are not recommended, acceptable results may be
obtained if the 1” fitting is a half coupling in formand the outer diameter of the coupling istightly
wrappedinvinyltapetodampenvibrations.