Users Manual

Immunity to
conducted
disturbances
induced by RF
fields
IEC 61000-4-6
3 V
0.15-80 MHz
6 V in ISM & Amateur
radio bands between
0.15 and 80 MHz
80% AM at 1kHz
Power supply line &
I/O lines
3 V
0.15-80 MHz
6 V in ISM &
Amateur radio bands
between 0.15 and 80
MHz
80% AM at 1kHz
The strength of RF field in the
frequency range higher than 150
kHz~80MHz, the strength of the
RF field is smaller than 3 V
Voltage dips,
short interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
0 % UT: 0.5 cycle
At 0°, 45°, 90°, 135°,
180°, 225°, 270° and
31
0 % UT; 1 cycle
and
70 % UT; 25/30
Cycles Single phase:
at 0º
0 % UT; 250/300
cycle
0 % UT: 0.5 cycle
At 0°, 45°, 90°, 135°,
180°, 225°, 270° and
31
0 % UT; 1 cycle
and
70 % UT; 25/30
Cycles Single phase:
at 0º
0 % UT; 250/300
cycle
The quality of supplied power is
suitable for general domestic or
home health care environment.
For the user to operate the
equipment continuously even the
electric power supply is
interrupted, it is recommended
that the uninterruptable power
supply device (UPS) or battery is
prepared.
Magnetic field of
supply frequency
(50/60Hz)
IEC 61000-4-8
30 A/m 30 A/m
If LED Lighting degradation of
performance occurs, it may be
necessary to position the Cybird
Curing Light further from sources
of power frequency magnetic
fields or to install magnetic
shielding. The power frequency
magnetic field should be
measured in the intended
installation location to assure that
it is sufficiently low.