Propaq® MD Service Manual September 2010 9650-0801-01 Rev.
ZOLL is a registered trademark, and SurePower is a trademark of ZOLL Medical Corporation. Propaq is a registered trademark of Welch Allyn, Inc. Nellcor is a trademark of Covidien. Oridion, Microstream, and FilterLine are registered trademarks of Oridion Systems, Ltd. All other trademarks and registered trademarks are property of their respective owners. Copyright © 2010 ZOLL Medical Corporation. All rights reserved.
Table of Contents Preface Overview................................................................................................................................ 1 Safety Considerations............................................................................................................ 2 Additional Reference Material................................................................................................ 2 Conventions ...........................................................................
TABLE OF CONTENTS Chapter 4 Life Cycle Counters Overview.............................................................................................................................. 69 Chapter 5 Replacement Parts Overview.............................................................................................................................. 71 Replacement Parts ..............................................................................................................
Preface Overview ZOLL Medical Corporation’s Propaq MD Service Manual is intended for the trained biomedical technician whose responsibility is to identify malfunctions and/or make repairs at the subassembly level. The Propaq MD Service Manual has six main sections and two appendixes. Preface—Contains safety warnings and an overview of the manual’s contents. Be sure to review this section thoroughly before attempting to use or service the Propaq MD unit.
PREFACE Chapter 6—Functional Description provides technical descriptions for the Propaq MD major subassembly modules. Appendix A—Propaq MD Electrical Hardware Interconnect Diagram and Propaq MD Signal and Power Interconnect Schematic. Appendix B—Contains simulators and settings that may be used to assess the performance of the NIBP module. Safety Considerations The following section describes general warnings and safety considerations for operators and patients.
Conventions Conventions WARNING! Warning statements describe conditions or actions that can result in personal injury or death. Caution Caution statements describe conditions or actions that can result in damage to the unit. Note: Notes contain additional information on using the defibrillator. Service Policy Warranty In North America: Consult your purchasing agreement for terms and conditions associated with your warranty. Outside of North America, consult ZOLL authorized representative.
PREFACE Technical Service for International Customers International customers should return the unit in its original container to the nearest authorized ZOLL Medical Corporation Service Center. To locate an authorized service center, contact the International Sales Department at ZOLL Medical at the above address. 4 9650-0801-01 Rev.
Chapter 1 Maintenance Tests Propaq MD Overview This chapter includes step-by-step procedures for the following types of maintenance tests: • • • • Calibration and Calibration Checks NIBP Valve, Pump, and Tubing Integrity Tests Operational Tests Functional Verification Tests At a minimum, a trained biomedical technician must perform a CO2 calibration and an NIBP calibration check once per year. A CO2 calibration check may be performed at any time to verify the CO2 calibration.
CHAPTER 1 MAINTENANCE TESTS This chapter describes the following maintenance tests: Calibration and Calibration Checks • NIBP Calibration Check • CO2 Calibration Procedure • CO2 Calibration Check NIBP Valve, Pump, and Tubing Integrity Tests • Leak Test • Deflation Test • Inflation Test Operational Tests • Physical Inspection of the Unit • Keypad Test • LED Test • LCD Test • RFU Indicator Test • Audio Test • Printer Test • Battery Test • Fan Test • NIBP Test Functional Verification Tests • Leads Test • Hear
Calibration and Calibration Checks Calibration and Calibration Checks NIBP Calibration Check Equipment Single lumen BP hose, a 500 cc volume, and a calibrated manometer or pressure gauge (< = 1mmHg accuracy). If a 500 cc volume is not available, an Adult BP cuff wrapped around a rigid cylinder (such as a glass jar) will suffice. Note: This test is to be performed for the following target pressures: 50 +/- 5 mmHg, 150 +/- 5 mmHg, and 250 +/- 5 mmHg. Procedure 1. Power on the Propaq MD. 2.
CHAPTER 1 MAINTENANCE TESTS Acceptable Results The unit passes the test if Pd-Primary is less than or equal to 3.0 mmHg and Pd-Safety is less than or equal to 4.0 mmHg for all three target pressures. Corrective Action If readings fail to meet the required accuracy, the NIBP module requires replacement or factory analysis and realignment. 8 9650-0801-01 Rev.
Calibration and Calibration Checks CO2 Calibration Procedure (required for annual calibration) Equipment 5% CO2 calibration gas cylinder with CO2 sample line (kit available from Scott Medical Products, PN 4653ORF-BD) Procedure 1. On the Propaq MD, do the following to enter the CO2 Calibration control panel: CO2 • Press the CO2 quick access key ( CO ). • Use the navigation keys to select Supervisor>Service>Diagnostic>CO2>Calibration. 2.
CHAPTER 1 MAINTENANCE TESTS CO2 Calibration Check Procedure Equipment 5% CO2 calibration gas cylinder with CO2 sample line (kit available from Scott Medical Products, PN 4653ORF-BD) Procedure 1. On the Propaq MD, do the following to enter the CO2 Calibration control panel: CO2 • Press the CO2 quick access key ( CO ). • Use the navigation keys to select Supervisor>Service>Diagnostic>CO2>Calibration. 2.
NIBP Valve, Pump, and Tubing Integrity Tests NIBP Valve, Pump, and Tubing Integrity Tests NIBP Leak Test Equipment Single lumen BP hose, a 500 cc volume, and a calibrated manometer. Note: If a 500 cc volume is not available, an Adult BP cuff wrapped around a rigid cylinder (such as a glass jar) will suffice Procedure 1. Power on the Propaq MD. 2. Connect the Propaq MD to the single lumen hose and the 100 cc volume.
CHAPTER 1 MAINTENANCE TESTS Corrective Action If the unit fails the test, the most likely cause is damaged or disconnected internal tubing. The device needs to be disassembled to replace or reattach the appropriate segment of NIBP tubing. The NIBP valves should also be inspected. NIBP Deflation Test Equipment Single lumen BP hose, a 500 cc volume, a timer, and a calibrated manometer (optional).
NIBP Valve, Pump, and Tubing Integrity Tests NIBP Inflation Test Equipment Single lumen BP hose, a 250 cc volume, a timer, and a calibrated manometer (optional). Procedure 1. Power on the Propaq MD. 2. Connect the Propaq MD to the single lumen hose and the 250 cc volume. If available, connect a calibrated manometer to the air line using a T-fitting. 3. In the Propaq MD, do the following to enter the NIBP Test control panel: 4. Press the Setup quick access key ( ). 5.
CHAPTER 1 MAINTENANCE TESTS Operational Tests The Propaq MD comes equipped with routine operational tests that can be performed periodically. With the exception of the Physical Inspection of the Unit test, these tests contain instructions within them provide guidance during the tests. Physical Inspection of the Unit Tools Needed None Test Setup None Observe this... Pass/Fail Housing 1. Is the unit clean and undamaged? o o 2. Does the unit show signs of excessive wear? o o 3.
Operational Tests Accessing the Operational Tests The operational tests on the unit can be accessed by doing the following: 1. Press the Setup quick access key ( ). 2. Use the navigation keys to select Supervisor>Service>Device Tests. The operational tests are displayed. Use the navigation keys to select one of the tests. 9650-0801-01 Rev.
CHAPTER 1 MAINTENANCE TESTS Keypad Test This test checks all the front panel buttons on the unit to make sure they are working correctly. Follow the instructions and use the navigation keys to move around the display and make selections. LED Test This test checks all the Light Emitting Diodes in the unit to make sure they are working correctly. Follow the instructions and use the navigation keys to move around the display and make selections. 16 9650-0801-01 Rev.
Operational Tests LCD Test This test checks the colors of the Liquid Crystal Display on the unit. Follow the instructions and use the navigation keys to move around the display and make selections. RFU Indicator Test This test checks the Ready For Use Indicator on the unit to make sure it is working correctly. Follow the instructions shown on the display. 9650-0801-01 Rev.
CHAPTER 1 MAINTENANCE TESTS Audio Test This test checks the audio tones on the unit to make sure they are working correctly. Follow the instructions and use the navigation keys to move around the display and make selections. Printer Test This test checks the printer to make sure it is working correctly. Follow the instructions shown on the display. 18 9650-0801-01 Rev.
Operational Tests Battery Test This test checks the battery to make sure it is working correctly. Follow the instructions on the display. Fan Test This test checks the fan in the unit to make sure it is working correctly. Follow the instructions and use the navigation keys to move around the display and make selections. 9650-0801-01 Rev.
CHAPTER 1 MAINTENANCE TESTS NIBP Test This allows the operator to perform various checks of the NIBP module to make sure it is working correctly. Follow the instructions in the NIBP Calibration Check procedure and other procedures outlined in this manual. Use the navigation keys to move around the display and make selections as required to execute the procedures. 20 9650-0801-01 Rev.
Functional Verification Tests Functional Verification Tests Before You Begin the Functional Verification Tests • Assemble the tools listed in following section, “Equipment You Need to Perform the • • • • • Functional Verification Tests.” Keep an extra fully charged Propaq MD battery available. Schedule an hour to conduct the entire maintenance test. Perform the tests in the order presented. Perform all the steps of each test procedure.
CHAPTER 1 MAINTENANCE TESTS Leads Test Tools Needed 3-lead, 5-lead, and 12-lead cables Test each cable separately. Test Setup Connect the lead wires appropriate to each test to the Fluke Impulse 7000DP Analyzer or equivalent. Do this... Observe this... Pass/Fail 1. Press the power switch to turn on the No ECG LEAD FAULT message Propaq MD. displayed. o o 2. Disconnect one lead from the simulator. The ECG LEAD FAULT message displays within 3 seconds (if configured). o o 3. Reconnect the lead.
Functional Verification Tests Heart Rate Display Test Tools Needed ECG simulator ECG Cable (3-lead, 5-lead or 12-lead) Test Setup Press the power switch to turn on the Propaq MD. Press Lead quick access key until II displays. Connect the ECG leads to the ECG simulator. Connect the ECG cable to the unit. 1. Do this... Observe this... Pass/Fail Set the ECG Simulator to 120BPM. The Heart Rate displays as 120 +/- 4 bpm o 9650-0801-01 Rev.
CHAPTER 1 MAINTENANCE TESTS Heart Rate Alarm Test Tools Needed ECG simulator Do this... Observe this... Pass/Fail 1. Press the power button on the Propaq MD to turn the unit on. Connect the ECG leads to the ECG simulator. Set the simulator to 120 BPM and press the Leads quick access key ( I,III...II, ) on the Propaq MD unit and select lead II. II displays next to the Leads quick access key. o o 2. On the Propaq MD: The Alarm Settings menu displays.
Functional Verification Tests Do this... Observe this... 10. On the Propaq MD, press the Alarm Suspend key ( ). Alarm is enabled. The alarm suspend key flashes. The alarm suspension timer disappears. o o 10. Set simulator to 160 BPM or higher. Heart Rate Value is highlighted and the alarm tone sounds. The HR/PR HIGH ALARM message displays. o o 11. On the Propaq MD, press the Alarm Silence/Reset button Alarm is suspended for 90 seconds. The alarm suspension timer appears in black.
CHAPTER 1 MAINTENANCE TESTS Leakage Current Test Tools Needed See the manufacturer’s instructions or supplied specifications for the leakage tester you use. Setup See the manufacturer’s instructions or supplied specifications for the leakage tester you use. Repeat leakage test with accessories: MFC, external paddles, internal paddles, and anterior/ posterior paddles. Perform these tests at the line-power voltage and frequency used in your installation.
Functional Verification Tests Shock Test Tools Needed Fluke Impulse 7000 DP Analyzer, or equivalent Test Setup None Do This... 1. Turn on the Analyzer. 2. On the Analyzer, select the DEFIB function. 3. On the Analyzer, select ENERGY. 4. Connect the Propaq MD MFE connector to the pads connection. 5. On the Propaq MD, apply power to the unit using a battery or an AC adapter. 6. On the Propaq MD, connect the MFE cable. 7. Press the power button to turn on the Propaq MD. 8.
CHAPTER 1 MAINTENANCE TESTS Synchronized Cardioversion Test Tools Needed Fluke Impulse 7000DP or equivalent defibrillator analyzer. Test Setup Connect the cable to the defibrillator analyzer. Select cardioversion on analyzer. Input 1mV ECG signal at 60 -120 BPM. Do this... 1. Press the power switch to turn on the Propaq MD, and press the Leads quick access key ( I,III...II, ) to select PADS. 2. Press the SYNC softkey on the defibrillator.
Functional Verification Tests Pacer Test Tools Needed Fluke Impulse 7000 DP Analyzer, or equivalent Test Setup None Do This... Observe This... 1. Turn on the Analyzer. 2. On the Analyzer, select the PACER function per test equipment instructions. 3. On the Analyzer, press F2 to select PACER JACKS. 4. Connect the MFC connector to the Analyzer. 5. On the Propaq MD, apply power to the unit using a battery or an AC adapter. 6. Press the power button to turn on the Propaq MD. 7.
CHAPTER 1 MAINTENANCE TESTS Printer Test Tools Needed None Test Setup None Do this... 1. 2. Observe this... Pass/Fail The recorded waveforms on the strip chart are clear and dark. o On the Propaq MD, install the paper tray on the left side of the unit. Press the Print quick access key ( the recorder. ) to start Note: If the printer fails to start, check setup for printer=yes. 3. 30 Press the Print quick access key ( ) to stop the recorder after approximately 10 seconds. o 9650-0801-01 Rev.
Functional Verification Tests SpO2 Test Tools Needed SpO2 Simulator: Fluke Index 2 or equivalent, NellcorTM SpO2 finger probe Test Setup Disconnect any ECG cable or IBP probes from the Propaq MD. Do This... Observe This... 1. Power on the Propaq MD without the SpO2 sensor attached. Verify that the SpO2 numeric displays INITIALIZING, changes to SEARCHING, then goes blank. 2. Connect the SpO2 sensor to the Propaq MD. Verify that SpO2 numeric displays SEARCHING. 3.
CHAPTER 1 MAINTENANCE TESTS EtCO2 Test Tools Needed CO2 sampling line airway adapter Test Setup None Do this... 1. 2. 3. 32 Observe this... Pass/Fail The CO2 trace is on the display. o o The CO2 waveform changes on the display. o o On Propaq MD, install the airway adapter set and sampling line in the connector at the side of the unit. Press the CO2 quick access key ( activate the CO2 trace. CO2 ) to On Propaq MD, gently blow into the airway adapter set while observing the display.
Functional Verification Tests IBP Test Tools Needed IBP patient simulator (with accuracy better than +/- 1 mmHg), or equivalent Setup Connect the test cables to IBP patient simulator. Do This... 1. Connect the IBP patient simulator to the Propaq P1 channel. 2. Set the IBP patient simulator to 0 mmHg. 3. On the Propaq MD, press > IBP. 4. On the Propaq MD, Press > P1->0<-. Observe This... Pass/Fail Display of 0/0(+/-2) in the P1 box at the bottom of the display (P1 is zeroed).
CHAPTER 1 MAINTENANCE TESTS NIBP Test Tools Needed NIBP simulator, single or dual lumen hose, adult, NIBP cuff, rigid cylinder Test Setup Connect the Propaq MD to the NIBP simulator using the single or dual lumen hose. Wrap the NIBP cuff around a rigid cylinder, such as a glass jar. Note: Refer to Appendix B for information on simulators and settings that you may use to assess the performance of the NIBP module. Do This... 34 1. On the Propaq MD, ensure the patient mode is set to Adult. 2.
Functional Verification Tests Temperature Test Tools Needed IBP Temp Simulator, or equivalent Test Setup None Do this... Observe This... 1. Connect the temperature simulator to the Propaq MD T1 channel. 2. On the IBP Temp Simulator, Select 98.6 The displayed temperature value is Degree F (36 Degree C) or a setting as close within 0.2 F or 0.1 C of the as possible as your simulator allows. simulator setting. 3. Repeat steps 1 and 2 of this procedure for the T2 channel. 9650-0801-01 Rev.
CHAPTER 1 MAINTENANCE TESTS 36 9650-0801-01 Rev.
Chapter 2 Troubleshooting Overview This chapter contains a list of error messages that users may see if the unit is not operating properly. If the errors you encounter are not listed below, call ZOLL Medical Corporation’s Technical Service Department for further assistance. (See “Technical Service” on page 3 for contact information.) Propaq MD Error Messages The following is a list of the Propaq MD error messages that may appear on your display.
CHAPTER 2 TROUBLESHOOTING 38 Error Message Explanation Technical Action QUICK ACCESS SW 1 SHORT Quick access key 1 is stuck. Replace the front panel. QUICK ACCESS SW 2 SHORT Quick access key 2 is stuck. Replace the front panel. QUICK ACCESS SW 3 SHORT Quick access key 3 is stuck. Replace the front panel. QUICK ACCESS SW 4 SHORT Quick access key 4 is stuck. Replace the front panel. QUICK ACCESS SW 5 SHORT Quick access key 5 is stuck. Replace the front panel.
Propaq MD Error Messages Error Message Explanation Technical Action PS DEFIB CHARGE SHORT Paddle set keys are stuck. Replace the paddle set or MFC. PS DEFIB RECORDER SHORT Paddle set keys are stuck. Replace the paddle set or MFC. PS DEFIB FIRE SHORT Paddle set keys are stuck. Replace the paddle set or MFC. PRINTER OFFLINE Printer module is off-line. If it remains off-line, try checking the connectors or power to the printer. If it persists, replace the printer module.
CHAPTER 2 TROUBLESHOOTING 40 Error Message Explanation Technical Action IBP DEVICE FAILURE IBP software detected a problem with the IBP data that is most likely a hardware issue. Replace the main board. IBP PROBE1 SHORTED An A/D value was detected that indicated the probe was shorted. Replace the IBP probe. If problem persists, replace the right side panel. IBP PROBE2 SHORTED An A/D value was detected that indicated the probe was shorted. Replace the IBP probe.
Propaq MD Error Messages Error Message Explanation Technical Action NIBP COMMUNICATION FAILURE There is a communication problem with the NIBP module. Ensure that the J32 connection is secure. If error persists, replace the NIBP module. NIBP CALIBRATION CHECK DUE Annual calibration check is due. Perform the NIBP Calibration Check procedure in Chapter 1, "Calibration and Calibration Checks". NIBP KINKED HOSE The cuff inflation was too rapid. There may be kinked tubing or another air obstruction.
CHAPTER 2 TROUBLESHOOTING 42 Error Message Explanation Technical Action NIBP AIR LEAK Major air leak or could not reach target pressure. Attempt to identify problems with the hose and/or cuff by doing the following: • Check the hose connection to the device. Make sure that the hose is tightly screwed onto the NIBP connector. • Ensure that the cuff is appropriately wrapped and is not too loose. • Inspect the cuff and hose for leaks. Inspect the O-ring on the NIBP hose connector.
Propaq MD Error Messages Error Message Explanation Technical Action NIBP NEONATAL DETECTED Device thinks a neonate cuff is connected. If no neonate cuff is connected, tubing may be damaged or kinked. Verify that neonate cuff is not connected. If problem persists, check the routing of the NIBP tubing inside the device. NIBP LOW VOLTAGE Insufficient voltage is supplied to the NIBP module. Check the module voltage in the NIBP Test screen. Inspect the unit or water ingress.
CHAPTER 2 TROUBLESHOOTING 44 Error Message Explanation Technical Action CO2 GAS INPUT OCCLUSION CO2 inlet blocked. Replace FilterLine®. If problem persists, check for CO2 inlet occlusion. If still persists, replace the CO2 module. CO2 MAINTENANCE REMINDER Periodic maintenance reminder. Replace CO2 module soon. CO2 MAINTENANCE REQUIRED Periodic Maintenance due. Replace CO2 module. CO2 NO FILTERLINE No filterline is detected. If FilterLine is connected and problem persists, replace FilterLine.
Propaq MD Error Messages Error Message Explanation Technical Action PS DEFIB ENERGY DOWN SHORT Paddle set energy down select button is stuck. Replace the paddles. PS DEFIB RECORDER SHORT Paddle set recorder button is stuck. Replace the paddles. DEFIB CAP LEAKAGE Capacitor bank cannot hold a charge. Replace the capbank module. DEFIB USING DEFAULT CALIBRATION Calibration data lost. Replace the defib board module. DEFIB FATAL CHARGE ERROR Charging error. Replace the defib board module.
CHAPTER 2 TROUBLESHOOTING 46 9650-0801-01 Rev.
Chapter 3 Disassembly Procedures Overview This chapter provides instructions on how to disassemble and reassemble the Propaq MD unit, and includes the following sections: • • • • • • • • • • • Required Equipment Safety Precautions Modules Handle Disassembly Printer Disassembly Main Unit Disassembly Front Enclosure Disassembly Display Disassembly NIBP and EtCO2 Disassembly Monitor Board and SpO2 Disassembly Defibrillator Board and Capacitor Bank Disassembly Required Equipment • • • • • • • • No.
CHAPTER 3 DISASSEMBLY PROCEDURES • • • • • • Kapton tape 3M copper adhesive tape, or equivalent 1/2” nut driver Large diagonal cutters Strong glue, such as Loctite 420 or equivalent Loctite needle tip dispenser Safety Precautions WARNING! Caution SHOCK HAZARD! TAKE THE NECESSARY PRECAUTIONS TO GUARD AGAINST SHOCK OR INJURY BEFORE YOU CONDUCT DEFIBRILLATOR TESTS OR REPAIRS. • • • • Only properly trained technicians should service the unit.
Removing the Handle Removing the Handle Tools Required • #1 Phillips screwdriver with torque capability • Orange (wooden) stick Setup • This procedure should be conducted on a static mat. • Wear a wrist strap to ground during the procedure. To reinstall the handle assembly, reverse the steps. Handle Disassembly 1 Using an orange stick, lift the screw covers off the handle in two locations. screw covers 2 Remove four (4) screws (2 on each handle) and remove the handle.
Removing the Printer Module Tools Required • #1 Phillips screwdriver with torque capability • Orange (wooden) stick Setup • This procedure should be conducted on a static mat. • Wear a wrist strap to ground during the procedure. To reinstall the printer module, reverse the steps. Printer Disassembly 1 Place the unit screen down on the mat in a horizontal position. 2 Open the printer housing and remove the printer paper. 3 Remove the two (2) screws inside the printer module in the illustrated areas.
Removing the Printer Module 6 Disconnect the two (2) cable connectors from the printer module. connectors 9650-0801-01 Rev.
Disassembling the Main Unit Tool Required • # 1 Phillips screwdriver with torque capability • orange (wooden) stick Setup • This procedure should be conducted on a static mat. • Wear a wrist strap to ground during the procedure. • Complete these procedures in the following order before performing this procedure: 1. “Removing the Handle” on page 49 2. “Removing the Printer Module” on page 50 To reassemble the main unit, reverse the steps.
Disassembling the Main Unit 6 Disconnect the battery communication cable (yellow, orange, brown, red) connector from the monitor board. battery cable 7 Disconnect the monitor/defibrillator flex cable from the monitor board. Note: Be careful to pull the connector straight out. Rocking the connector or disconnecting it at an angle may result in bent pins. 8 Disconnect the three power cable connectors (battery (black), AUX (white), and defibrillator (red)).
11 54 Separate the two parts of the unit. 9650-0801-01 Rev.
Removing the Front Enclosure Module Removing the Front Enclosure Module Tools Required • #2 Phillips screwdriver with torque capability • Orange (wooden) stick Setup • This procedure should be conducted on a static mat. • Wear a wrist strap to ground during the procedure. • Complete these procedures in the following order before performing this procedure: 1. “Removing the Handle” on page 49 2. “Disassembling the Main Unit” on page 52 To reinstall the front enclosure module, reverse the steps.
6 56 Remove the ZIF (Zero Insertion Force) flex cable using an orange stick. 9650-0801-01 Rev.
Removing the Display Module Removing the Display Module Tools Required • #1 Phillips screwdriver with torque capability • 3M copper adhesive tape • 6mm socket wrench Setup • This procedure should be conducted on a static mat. • Wear a wrist strap to ground during the procedure. • Complete these procedures in the following order before performing this procedure: 1. “Disassembling the Main Unit” on page 52 2. “Removing the Front Enclosure Module” on page 55 To reinstall the display module, reverse the steps.
5 Flip the unit back over so the display screen is facing up. 6 Lift the casting up while pushing down on the rubber boot that runs around the edge of the LCD screen. boot 7 Remove the copper tape from the casting. Note: During reassembly, apply new copper tape for grounding. copper tape 8 Disconnect the LCD display cable connectors from the back of the LCD screen to release it. cable connectors 9 Remove the Kapton tape from the display cable. Kapton tape 58 9650-0801-01 Rev.
Removing the EtCO2 and NIBP Modules Removing the EtCO2 and NIBP Modules Tool Required • # 1 Phillips screwdriver with torque capability • orange (wooden) stick Setup • This procedure should be conducted on a static mat. • Wear a wrist strap to ground during the procedure. • Complete these procedures in the following order before performing this procedure: 1. “Removing the Handle” on page 49 2. “Removing the Printer Module” on page 50 3. “Disassembling the Main Unit” on page 52 4.
3 Disconnect multi-wire IBP cable with ferrite in the area illustrated. 4 Cut the tie wrap to the IBP cable in the area shown with the orange stick. multi-wire IBP cable Note: During reassembly, secure a new tie wrap to the IBP cable. 5 Disconnect the USB flex cable from ZIF connector in the area illustrated. 6 Lay front end of separated unit down with opened side facing up. 7 Disconnect the black EtCO2 tube connector from the air vent.
Removing the EtCO2 and NIBP Modules 9 Remove five (5) screws and washers from the NIBP unit to release it. Note: During reassembly, torque screws to 6 in-lbs. 10 Remove multi-wire EtCO2 cable with ferrite. 11 Disconnect the brown SpO2 flex cable as shown. 12 Disconnect the NIBP tubing from the side panel in the area illustrated. 13 Remove the NIBP module from the unit. 9650-0801-01 Rev.
14 Tilt side panel to access the EtCO2 connector and use an orange stick release the tabs and remove the EtCO2 connector. EtCo EtCO2 connector after removal 2 connector after removal EtCo2 connector before removal 15 Remove the EtCO2 module from the unit. 16 Remove three (3) screws from the outside of the left side panel in the areas illustrated. Note: During reassembly, torque screws to 6 in-lbs.
Removing the Monitor Board and SpO2 Modules Removing the Monitor Board and SpO2 Modules Tool Required • # 1 Phillips screwdriver with torque capability Setup • This procedure should be conducted on a static mat. • Wear a wrist strap to ground during the procedure. • Complete these procedures in the following order before performing this procedure: 1. “Removing the Handle” on page 49 2. “Removing the Printer Module” on page 50 3. “Disassembling the Main Unit” on page 52 4.
3 Lift the monitor board to expose the multi-wire LCD cable on the back side of monitor board and disconnect it. 4 Remove three (3) screws and washers from the NIBP unit. Note: During reassembly, torque screws to 6 in-lbs. 64 5 Lift the SpO2 unit straight up to disconnect it from the monitor board and remove it from the unit. 6 Lift out the monitor board. 9650-0801-01 Rev.
Removing the Defibrillator Board and Capacitor Bank Modules Removing the Defibrillator Board and Capacitor Bank Modules Tools Required • #1 Phillips screwdriver with torque capability • #2 Phillips screwdriver with torque capability • 26 mm open end wrench Setup • This procedure should be conducted on a static mat. • Wear a wrist strap to ground during the procedure. • Complete these procedures in the following order before performing this procedure: 1. “Removing the Handle” on page 49 2.
5 Remove screw in the middle of the plastic isolator (located on top of the processor board) in the area illustrated. Note: During reassembly, torque screw to 6 in-lbs. 6 Remove the plastic isolator shield. 7 Remove the four (4) screws in the areas illustrated. Note: During reassembly, torque screws to 6 in-lbs. 8 66 Lift up the processor board and remove the brown flex cable (there are two connections). 9 Lift out the processor board.
Removing the Defibrillator Board and Capacitor Bank Modules 11 Locate the black ferrite. Using the orange stick, pull the ferrite cover apart and untwist the couple. ferrite 12 13 Cut the tie wrap securing the black and read coupled wire (in illustrated area) to remove the power cell. Disconnect the black and red wire (via the red connector) in the area illustrated. tie wrap power cell red connector 14 Remove four (4) screws in the areas illustrated.
68 9650-0801-01 Rev.
Chapter 4 Life Cycle Counters Overview This chapter contains instructions for setting life cycle counters on the Propaq MD. The Propaq MD tracks various life cycle counters for specific modules installed in the device. When replacing modules in the unit, it is important to reset the life cycle counters.
CHAPTER 4 LIFE CYCLE COUNTERS CO2 On the Propaq MD, do the following: 1. Press the Setup quick access key ( ). 2. Use the navigation keys to select Supervisor>Service>Diagnostic>CO2>Service Codes. 3. Select Reset CRC Counter. Select Yes when prompted to confirm the reset. See “Functional Verification Tests” on page 1-21 to perform the functional verification tests. SpO2 On the Propaq MD, do the following: 1. Press the Setup quick access key ( ). 2.
Chapter 5 Replacement Parts Overview This section contains a listing of the replacement parts available for the ZOLL Propaq MD devices. Replacement parts may be ordered through an authorized ZOLL distributor or directly from ZOLL Medical Corporation. The prices for parts are available from ZOLL Medical Corporation’s Technical Service Department.
CHAPTER 5 REPLACEMENT PARTS Replacement Parts 72 PART NUMBER QUANTITY DESCRIPTION 705519 1 Handle Top 705517 1 Handle Bottom 705510 4 Screw 0162-0263 4 Screw Mount 705523 1 Left Cap, Handle 705524 1 Right Cap, Handle 1017-0005-01 1 Front Panel Assembly 403889 1 Panel I/O Left 402359 1 Panel I/O Right 705099-99 1 Back Housing 708494-99 1 Middle Housing 9301-0225 1 Dock Connector Assembly 402342 1 Cable Harness, Printer 9500-0650 1 Assy, Recerpt Defib Cable 4038
Replacement Parts PART NUMBER QUANTITY DESCRIPTION 7772-0001-01 1 Printer Assembly 7772-0002-01 1 Front Enclosure Assembly 7772-0003-01 1 Display Assembly 7772-0004-01 1 NIBP Assembly 7772-0005-01 1 Oridion Assembly 7772-0006-01 1 Monitor Board Assembly 7772-0007-01 1 CapBank Assembly 7772-0008-01 1 Defib Board Assembly 7772-0009-01 1 SpO2 Assembly 9650-0801-01 Rev.
CHAPTER 5 REPLACEMENT PARTS 74 9650-0801-01 Rev.
Chapter 6 Functional Description Overview This chapter contains functional descriptions of the components contained in the Propaq MD and the Propaq MD options. Circuit Distribution The Defib/Monitor Propaq MD unit consists of two key PCB assemblies, the Monitor Board and the Defibrillator Board as well as a few modules and assemblies: the Front Panel Keypad assembly, Status LCD Board, Dock Connector Board, Display, and Printer.
CHAPTER 6 FUNCTIONAL DESCRIPTION Main Processor (MP) The MP section controls the user interface, connectivity needs, as well as data logging, printing, and general system control. It consists of the CPU, program memory Flash and SDRAM, Communication with modules/circuits: USB Client, USB Host, Debug Ports (JTAG,Serial), Storage memory for 24 hour disclosure data, Internal VGA display and interface circuit, LED indicators, Key Pad, and Audio circuit.
Overview isolated serial bus through the FPGA. The IBP probes connect to the monitor board through the IBP/Temp wire harness which connects to J33 of the Monitor board. SPO2 Module Propaq MD supports the Nellcor NELL-3 OEM Module that mounts to the Monitor board and communicates through J36 and J40. The SpO2 module communicates across an isolation barrier using a UART on the PP. Signals from the probe are routed to the SpO2 module through a flex cable that connects to J41 of the Monitor board.
CHAPTER 6 FUNCTIONAL DESCRIPTION module contains power supplies, charging and discharge circuitry, pacing circuitry, and the Pads/Paddles ECG preamplifier and impedance circuitry. Communication with the Defibrillator/Pacer module is a function of the PP. The ECG-Charger board contains the charger circuit, ECG acquisition circuitry, and protection components. The Processor-Bridge-Pacer board contains the micro controller, CPLD, Pacer, and defibrillation shock delivery circuit.
Overview Printer Propaq MD uses an internal thermal printer of 50 mm paper size that is built into the Defib section of the enclosure. Communications with the printer is bi-directional, serial, (Main CPU to/from printer) and complies with the software/hardware interface of the internal printer. Communications are implemented using hardware Universal Asynchronous Receiver. Transmitter (UART). The printer is controlled by the MP and is connected through a cable to J46 and J49 of the Monitor board.
CHAPTER 6 FUNCTIONAL DESCRIPTION 80 9650-0801-01 Rev.
Appendix A Overview This appendix includes the following diagrams: • Propaq MD Electrical Hardware Interconnect Diagram • Propaq MD Signal and Power Interconnect Schematic 9650-0801-01 Rev.
Propaq MD Electrical Hardware Interconnect Diagram 82 9650-0801-01 Rev.
Propaq MD Signal and Power Interconnect Schematic Propaq MD Signal and Power Interconnect Schematic 9650-0801-01 Rev.
84 9650-0801-01 Rev.
Appendix B Overview This appendix includes simulators and settings that may be used to assess the performance of the NIBP module. Readings should be obtained within +/- 10 mmHg of the expected measurement result. The following simulators are used in this chapter: • • • • Bio-Tek Instruments BP Pump Bio-Tek Instruments BP Pump2 DNI Nevada CuffLink Clinical Dynamics Smart Arm NIBP Simulator Performance FastBP A dual lumen hose is required for FastBP.
Bio-Tek Instruments BP Pump Adult/Pediatric Mode FastBP StepBP Simulator Setting Expected Systolic Expected Diastolic Expected Systolic Expected Diastolic 200/150 198 151 200 149 120/80 120 83 116 84 80/50 83 54 77 53 Neonate Mode NeoBP Simulator Setting Expected Systolic Expected Diastolic 60/30 56 26 35/15 31 12 Bio-Tek Instruments BP Pump2 Adult/Pediatric Mode FastBP StepBP Simulator Setting Expected Systolic Expected Diastolic Expected Systolic Expected Diastolic 20
Overview DNI Nevada CuffLink Adult/Pediatric Mode FastBP StepBP Simulator Setting Expected Systolic Expected Diastolic Expected Systolic Expected Diastolic 200/150 210 152 201 138 120/80 116 77 124 80 80/50 87 49 90 50 Clinical Dynamics Smart Arm Adult/Pediatric Mode FastBP StepBP Simulator Setting Expected Systolic Expected Diastolic Expected Systolic Expected Diastolic 200/150 192 152 194 136 120/80 124 79 122 80 80/50 85 42 80 49 Neonate Mode NeoBP Simulator S
88 9650-0801-01 Rev.