Acceptable Results
The unit passes the test if P
d-Primary
 is less than or equal to 3.0 mmHg and P
d-Safety
 is less than 
or equal to 4.0 mmHg for all three target pressures.
Corrective Action
If readings fail to meet the required accuracy, the NIBP module requires replacement or factory 
analysis and realignment.