Service manual

Functional Verification Tests
9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 27
Shock Test
Tools Needed Fluke Impulse 7000 DP Analyzer, or equivalent
Test Setup None
Table 2. Delivered Energy at 200J Defibrillator Setting into a Range of Loads
Note: For a complete listing of Propaq MD Delivered Energy at Every Defibrillator Setting into a
Range of Loads
, see Appendix A in the Propaq MD Operators Guide.
Do This... Observe This... Pass/Fail
1. Turn on the Analyzer.
2. On the Analyzer, select the DEFIB function.
3. On the Analyzer, select ENERGY.
4. Connect the Propaq MD MFE connector to
the pads connection.
5. On the Propaq MD, apply power to the unit
using a battery or an AC adapter.
6. On the Propaq MD, connect the MFE cable.
7. Press the power button to turn on the Propaq
MD.
8. On the Propaq MD, set energy level to 200J
by pressing the Select Energy UP arrow.
9. On the Propaq MD, press the CHARGE
button.
The unit charges to 200J, the red
Shock button lights up and a
ready tone is heard.
oo
10. On the Propaq MD, press the SHOCK
button.
11. On the Analyzer, observe and record the
energy measurement.
Energy is delivered at 200J.
Verify that the delivered energy is
within 15% based on the values in
the following table.
oo
Selected
Energy
Load
25 50 75 100 125 150 175 Accuracy*
200 J 142 J 230 J 249 J 253 J 269 J 261 J 260 J ±15%