Service manual

Functional Verification Tests
9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 29
Pacer Test
Tools Needed Fluke Impulse 7000 DP Analyzer, or equivalent
Test Setup None
Do This... Observe This... Pass/Fail
1. Turn on the Analyzer.
2. On the Analyzer, select the PACER
function per test equipment instructions.
3. On the Analyzer, press F2 to select PACER
JACKS.
4. Connect the MFC connector to the
Analyzer.
5. On the Propaq MD, apply power to the unit
using a battery or an AC adapter.
6. Press the power button to turn on the
Propaq MD.
7. On the Propaq MD, press the PACER
button.
8. On the Propaq MD, use the navigation keys
to do the following:
•Set Rate to 30.
•Set Output to 10.
•Select Start Pacer.
•Pacer Rate PPM
(28.5ppm-31.5ppm)
•Pacer amplitude output
(5mA-15mA)
•Pacer pulse width (38ms to 42ms)
oo
9. On the Impulse 7000 DP Analyzer, record
the following:
•pulse rate
•pacer amplitude output
•pacer pulse width
10. On the Propaq MD, use the navigation keys
to do the following:
•Set Rate to 180.
•Set Output to 140.
•Select Start Pacer.
•Pacer Rate PPM (162ppm-198ppm)
•Pacer amplitude output
(133mA-147mA)
•Pacer pulse width (38ms to 42ms)
oo
11. On the Impulse 7000 DP Analyzer, record
the following:
•pulse rate
•pacer amplitude output
•pacer pulse width